Preparing for Use
R&S
®
NRP-Zxx
16
Getting Started 1176.8822.02 ─ 04
To connect to an Rohde & Schwarz instrument
► Insert the host connector, red marking upwards, into the SENSOR connector
of your instrument.
4.5
Disconnecting the R&S NRP
‑
Z Power Sensors
Additional requirements must be fulfilled when disconnecting the R&S
NRP-
Z27/-Z28/-Z37/-Z98 sensors. Refer to the operating manual of the corre-
sponding sensor for a description on how to disconnect it.
4.5.1
Disconnecting the RF Connector of the R&S NRP
‑
Z Power
Sensors
Risk of damage to the center pin of the RF connector
When disconnecting the power sensor, always rotate only the union nut of
the RF connector. Do not rotate the sensor itself to avoid damage the cen-
ter pin of the connector.
► Carefully loosen the screw at the front of the sensor's RF connector and
remove the sensor.
4.5.2
Disconnecting the Host Connector of the R&S NRP
‑
Z
Power Sensors
To disconnect from an R&S
NRP
‑
Z3 or R&S
NRP
‑
Z4 adapter
► Pull at the host connector (
) ) until it disconnects from the con-
nector of the USB adapter.
Disconnecting the R&S NRP
‑
Z Power Sensors