Remote Control Commands
R&S
®
NRPxxTWG
64
User Manual 1178.8371.02 ─ 04
Example:
FREQ 10000
7.8.3
Configuring Corrections
It is possible to set some parameters that compensate for a change of the measured
signal due to fixed external influences.
........................................................................................... 64
.......................................................................................... 65
............................................................................................ 67
.................................................................................. 69
7.8.3.1
Duty Cycle Corrections
The duty cycle is the percentage of one period during which the signal is active, when
pulse-modulated signals are corrected. The duty cycle is only evaluated in the Continu-
ous Average mode.
Remote commands:
[SENSe<Sensor>:]CORRection:DCYCle
............................................................................ 64
[SENSe<Sensor>:]CORRection:DCYCle:STATe
.................................................................. 64
[SENSe<Sensor>:]CORRection:DCYCle
<duty_cycle>
Available in continuous average mode.
Sets the duty cycle for measuring pulse-modulated signals. The duty cycle defines the
percentage of one period during which the signal is active. If the duty cycle is enabled,
the R&S NRPxxTWG takes this percentage into account when calculating the signal
pulse power from the average power.
Parameters:
<duty_cycle>
Range:
0.001 to 100.00
*RST:
1.00
Default unit: Percent
[SENSe<Sensor>:]CORRection:DCYCle:STATe
<state>
Enables or disables the duty cycle correction for the measured value.
Parameters:
<state>
*RST:
OFF
7.8.3.2
Offset Corrections
The offset accounts for external losses by adding a fixed level offset in dB.
The attenuation of an attenuator located ahead of the sensor (or the coupling attenua-
tion of a directional coupler) is taken into account with a positive offset, i.e. the sensor
Configuring Basic Measurement Parameters