Remote Control Commands
R&S
®
NRPxxTWG
45
User Manual 1178.8371.02 ─ 04
Trigger source
Description
Remote commands to initiate the measurement
"External
2"
Uses the digital input signal supplied at the SMB
connector.
"Bus"
Triggered by the remote command.
7.5.2.4
Dropout Time
The dropout time is useful when dealing with signals with several active slots, for
example GSM signals, see
. When measuring in sync with the signal, a trig-
ger event is to be produced at A, but not at B or C.
Figure 7-1: Significance of the dropout time
The RF power between the slots is below the threshold defined by the trigger level and
the trigger hysteresis. Therefore, the trigger hysteresis alone cannot prevent triggering
at B or at C. Therefore, set the dropout time greater than the time elapsed between
points D and B and between E and C, but smaller than the time elapsed between F
and A. Thus, you ensure that triggering takes place at A.
Because the mechanism associated with the dropout time is reactivated whenever the
trigger threshold is crossed, you can obtain also unambiguous triggering for many
complex signals.
If you use a hold-off time instead of a dropout time, you can obtain stable triggering
conditions - regular triggering at the same point. But you cannot achieve exclusive trig-
gering at A.
7.5.2.5
Hold-Off Time
During the hold-off time, a period after a trigger event, all trigger events are ignored.
Controlling the Measurement