Remote Control
R&S
®
ESRP
920
User Manual 1175.7445.02 ─ 08
Example:
See
"Designing a transducer factor"
Manual operation:
See
"Defining characteristics of a transducer factor"
[SENSe:]CORRection:TRANsducer:DATA
<Frequency>,<Level>
This command defines the shape of the transducer factor.
A transducer factor may contain up to 625 reference values. Each reference value is a
combination of a frequency and a level value.
Before you can use this command, you have to select a transducer factor with
[SENSe:]CORRection:TRANsducer:SELect
Parameters:
<Frequency>
Numeric value in Hz.
The frequency values have to be in ascending order.
<Level>
The unit depends on
[SENSe:]CORRection:TRANsducer:
on page 921. By default, it is dB.
Example:
See
"Designing a transducer factor"
Manual operation:
See
"Defining characteristics of a transducer factor"
[SENSe:]CORRection:TRANsducer:DELete
This command deletes a transducer factor.
Before you can use the command, you have to select a transducer factor with
[SENSe:]CORRection:TRANsducer:SELect
.
Example:
CORR:TRAN:SEL 'Transducer 1'
CORR:TRAN:DEL
Deletes the selected transducer.
Usage:
Event
Manual operation:
See
"Defining characteristics of a transducer factor"
[SENSe:]CORRection:TRANsducer:SCALing
<Scale>
This command selects the scale of the transducer factor.
Before you can use the command, you have to select a transducer factor with
[SENSe:]CORRection:TRANsducer:SELect
ESRP.
Parameters:
<Scale>
LINear | LOGarithmic
*RST:
LINear
Example:
See
"Designing a transducer factor"
Using Transducers