Quick Start
R&S
®
ESRP
116
User Manual 1175.7445.02 ─ 08
OFF] key on the front panel; see
Chapter 3.2.1.7, "Switching the Instrument On and
on page 41. If the instrument is preset, it loads the presettings.
You can alter these settings and define a settings file to be loaded. This requires per-
forming the following procedure. Be aware that the chosen settings file is loaded both
while booting and presetting.
1. Press the [SAVE/RCL] key.
2. Press the "Startup Recall" softkey.
3. Press the "Startup Recall (On/Off)" softkey to activate the recall function.
4. Press the "Select Dataset" softkey.
The "Startup Recall" dialog box is displayed.
5. If necessary, select the path under which the file to be loaded is stored.
6. Select the settings file to be loaded (DFL file).
7. Press the "Select" button in the dialog box.
Basic Measurement Examples