Preparing for use
R&S
®
AMN6500
19
User Manual 1352.1213.02 ─ 01
1
4.8
Measurement setup
The R&S
AMN6500 Two Line V-Network is suitable for measurments against CISPR,
EN, VDE, ANSI, FCC Part 15 and MIL-STD-461 standards.
The measurement setups prescribed in the individual standards are largely similar. A
detailed description is provided for example by the standard CISPR 14-1 and EN
55014-1, see also the following figure.
Devices under test that are not equipped with permanently attached connecting wires
are connected to the network using a 1 meter long, unshielded wire. For devices under
test equipped with a permanently attached cable that is longer than 1 meter, the cable
is laid out in a folded configuration.
Measurement setup