![Rohde & Schwarz 1352.1313.12 User Manual Download Page 17](http://html1.mh-extra.com/html/rohde-and-schwarz/1352-1313-12/1352-1313-12_user-manual_1477785017.webp)
Preparing for use
R&S
®
AMN6500
17
User Manual 1352.1213.02 ─ 01
DIN/EN/IEC/UL/CSA 62368-1. An external power supply without reinforced/double
insulation is live in case of failure. Touching a live electrical device causes serious
personal injury, or even death.
Connect the external power supply to the "EXT. POWER SUPPLY" socket (1).
The input is reverse voltage protected.
1
4.5
Connecting to reference ground
The network is connected to the PE grounding connection for the mains power supply
through the mains cable, but this is not adequate for use as a reference ground in HF
noise voltage measurement.
The reference ground used can be attached flat to the grounding rail (2) on the side of
the device by a broad sheet metal with three metric M4 screws (1).
1
1
1
2
1 = Metric M4 screw
2 = Grounding rail
Connecting to reference ground