Rohde & Schwarz 1328.6660K02 Getting Started Download Page 13

Preparing for use

R&S

®

FPC

13

Getting Started 1328.7409.02 ─ 05

To unmount the R&S

 

FPC from a rack

1. Loosen the screws at the rack brackets.

2. Remove the R&S

 

FPC from the rack.

3. If placing the R&S

 

FPC on a bench top again, unmount the adapter kit from

the R&S

 

FPC. Follow the instructions provided with the adapter kit.

3.5

 

Considerations for test setup

Signal input and output

An unsuitable test setup can damage the instrument and connected devices.
Before you switch on the instrument, check that all inputs and outputs are cor-
rectly connected and that the signal levels are within the specified ranges defined
in the data sheet.

In particular, pay attention to the following limits to avoid damage to the
R&S

 

FPC:

Do not overload the 

RF input

 and keep within the maximum allowed signal

level defined in the data sheet.

Never exceed a DC input voltage of 50

 

V at the 

RF input

.

Never exceed a reverse power of 23

 

dBm at the 

signal source output

.

Cable selection and electromagnetic interference (EMI)

Electromagnetic interference (EMI) can affect the measurement results.

To suppress electromagnetic radiation during operation:

Use high-quality shielded cables, for example, double-shielded RF and LAN
cables.

Always terminate open cable ends.

Ensure that connected external devices comply with EMC regulations.

Considerations for test setup

Summary of Contents for 1328.6660K02

Page 1: ...R S FPC Spectrum Analyzer Getting Started 1328740902 Version 05 L 92...

Page 2: ...de schwarz com Internet www rohde schwarz com Subject to change data without tolerance limits is not binding R S is a registered trademark of Rohde Schwarz GmbH Co KG Trade names are trademarks of the...

Page 3: ...e information is provided in many languages as printed Safety Instructions The printed Safety Instructions are delivered with the product Throughout the documentation safety instructions are provided...

Page 4: ...d you can operate the product up to an altitude of 2000 m above sea level The product is suitable for pollution degree 2 environ ments where nonconductive contamination can occur For more information...

Page 5: ...ularly to ensure that they are undamaged Also ensure that nobody can trip over loose cables If the product needs an external power supply use the power supply that is delivered with the product or tha...

Page 6: ...the casing inform about Personal safety see Meaning of safety labels on page 6 Product and environment safety see Table 1 1 Identification of the product see Device ID on page 19 Table 1 1 Labels rega...

Page 7: ...ts with a serial number 200000 have an internal Wi Fi module You can access the regulatory information for this module in the firmware of the R S FPC 1 Press the Setup key The R S FPC opens the Instru...

Page 8: ...S FPC user manual there is a separate user manual for the R S InstrumentView software package This manual contains a description of all features of the R S InstrumentView software package The online...

Page 9: ...ense texts of the used open source software www rohde schwarz com manual fpc The open source acknowledgement is also integrated into the firmware chm for mat You can export the file to a memory stick...

Page 10: ...13 Connecting AC power 14 Turning the R S FPC on and off 14 3 1 Lifting and carrying See Lifting and carrying the product on page 4 3 2 Unpacking and checking the instrument 1 Unpack the R S FPC caref...

Page 11: ...en in the data sheet under Gen eral data Class B equipment is suitable for use in Residential environments Environments that are directly connected to a low voltage supply network that supplies reside...

Page 12: ...To prepare the rack 1 Observe the requirements and instructions in Setting up the product on page 4 2 NOTICE Insufficient airflow can cause overheating and damage the product Design and implement an e...

Page 13: ...levels are within the specified ranges defined in the data sheet In particular pay attention to the following limits to avoid damage to the R S FPC Do not overload the RF input and keep within the max...

Page 14: ...afety class EN61010 Therefore it may only be connected to an outlet that has a ground contact The AC power supply has the following characteristics Line voltage 100 V AC to 240 V AC Line frequency 50...

Page 15: ...er switch on the rear panel of the R S FPC position O The instrument is no longer supplied with AC power To disconnect from power The product is in the standby state 1 NOTICE Risk of data loss If you...

Page 16: ...panel of the R S FPC 1 USB ports type A 2 Headphone jack 3 Softkeys 4 Signal source output 5 Function keys and alphanumeric keypad 6 RF Input 7 Power switch Power switch The power switch turns the R S...

Page 17: ...0 ports on the front panel type A allow you to connect devices like memory sticks Function keys and alphanumeric keypad The function keys provide access to the measurement settings and functions The...

Page 18: ...ce 2 LAN 3 USB port type B 4 Power supply Power supply The AC power supply and main power switch are located in a unit on the rear panel of the instrument The main power switch has the following state...

Page 19: ...printouts or data transfer The assignment of the RJ 45 connector supports twisted pair category 5 UTP STP cables in a star configuration UTP stands for unshielded twisted pair and STP for shielded twi...

Reviews: