108
On completion of cleaning:
3
With the cover off, let the test strip guide
dry for
at least 10 minutes
before re-
attaching the test strip guide cover and
testing again.
•
Visually verify that no residual moisture is
seen anywhere on the test strip guide and
cover at the completion of cleaning.
•
Ensure that the test strip guide and cover
are completely dry before assembling
them.
WARNING
4
Re-attach the test strip guide cover to
the housing. Make sure that the cover is
properly closed. You will hear it snap into
place.
Cleaning and Disinfecting the Meter and the Lancing Device
CoaguChek Vantus User’s Manual, version 1.0