136
Cedex HiRes Analyzer Operator’s Guide
E
E
8.1.2.
Running a Water Blank
Before an
SST including Measurement
can be run, a water blank (
Water For Injection
[WFI]) should be run to
check the condition of the reagents. No more than 20 counts should be recorded for this measurement. This can
be scheduled using the
SST including Measurement
button (see “
Creating SST checkpoints with and without
measurements
”). The WFI measurement requires a relative deviation rather than an absolute deviation, which
means that 0 cannot be used as a
Target Value
as a relative deviation from 0 will always be 0.
Therefore, when setting up a WFI measurement, the
Target Value
should be set to 10 with a relative deviation
of 100%. This will result in an acceptance range of 0 – 20 objects (see Table 2).
8.1.3.
Setting up the SSTFailureReaction in the System Options Menu
The reaction of the system to an SST failure depends on the confi guration in the
System Options
. The
System
Options
can be accessed under the
Setup
menu on the Control Center:
1
Open the
System Options
and click on the
General
tab.
2
Look for the parameter
SSTFailureReaction
.
Entering
Restrict
in the
Value
column will lead to prohibition of any further measurements, including
SST measurements, in the event of an SST failure. This will require a manual unlocking of the system in
the
Lock – Unlock System
area before measurement activity can be resumed (see “
Locking – Unlocking
the System
”).
If
Warning
is entered, measurement activity can be continued but there will be a warning about SST
failure before a measurement is carried out.
Choosing the
Audit
option will simply lead to a log entry in the audit trail fi le in the case of an SST failure.
There will not be any restrictions or warnings.
3
Click on
Save
to store the change.
■
Measurements that are run despite the fact that not all SST procedures are up to date or valid will have no
check-mark in the
SST Valid
box in the
Measurement
window.
The System Suitability Test (SST)
Creating SST Checkpoints
Summary of Contents for Cedex HiRes
Page 8: ......
Page 14: ...8 Cedex HiRes Analyzer Operator s Guide...
Page 24: ...18 Cedex HiRes Analyzer Operator s Guide...
Page 164: ...158 Cedex HiRes Analyzer Operator s Guide G G Index...
Page 165: ......