![Roche Accu-Chek Rapid-D Link Instructions For Use Manual Download Page 11](http://html1.mh-extra.com/html/roche/accu-chek-rapid-d-link/accu-chek-rapid-d-link_instructions-for-use-manual_1472328011.webp)
9
en
Fill volumes of the complete set:
30 cm/12”
approx. 7 units U 100 insulin
approx. 70 μl
60 cm/24”
approx. 10 units U 100 insulin
approx. 100 μl
80 cm/31”
approx. 13 units U 100 insulin
approx. 130 μl
110 cm/43”
approx. 18 units U 100 insulin
approx. 180 μl
Picture 4
Hold the Accu-Chek Rapid-D Link firmly between thumb and forefinger at the disc holder
(E). With the other hand, remove the adhesive backing (D) from the self-adhesive (C) and
twist the protective cover (B) off the steel cannula (A).
Picture 5
With your free hand, pinch together a bit of subcutaneous tissue at the disinfected site.
Insert the Accu-Chek Rapid-D Link under the skin at a perpendicular angle. Press the self-
adhesive (C) firmly onto the skin in order to ensure secure seating of Accu-Chek Rapid-D
Link. Bend the disc holder (E) slightly sideways and remove it. In doing this, make sure that
the self-adhesive comes to lay flat on the skin.
WARNING
Change the head-set every 2 days (max. 48 hours). Otherwise you risk insulin resistance
or inflammation of the infusion site.
Never use the transfer-set for more than 6 days.
Picture 6
Form a loop to relieve stress and secure it into place with a strip of adhesive plaster. Start
distribution through the pump.