224
NOTE
j
If data is not available, the
No Data Available
screen is displayed.
j
Trend graphs do not include corrupt results or control results.
j
On the upper right side of the graph, the maximum bolus amount indicator is the pink tick mark with the largest value shown
beside it (on this example screen: 15 U). The purpose of this indicator is to adjust the scale for the top portion of the graph
based on the single biggest bolus insulin amount (callout 11) delivered for the time chosen (callout 6). The maximum bolus
amount indicator values available on the meter are 1, 5, 15, 30 and 60 U. For example, if the single biggest bolus insulin
amount is 8 U, then the meter scales the upper graph to be between 0 and 15 U.
73627_07030355001_F_Aviva_EN.indb 224
2/22/2018 10:10:02 AM
Summary of Contents for Accu-Chek Aviva Insight
Page 2: ...73627_07030355001_F_Aviva_EN indb 2 2 22 2018 10 08 01 AM ...
Page 52: ...46 73627_07030355001_F_Aviva_EN indb 46 2 22 2018 10 08 20 AM ...
Page 78: ...72 73627_07030355001_F_Aviva_EN indb 72 2 22 2018 10 08 31 AM ...
Page 104: ...98 Select Save 9 73627_07030355001_F_Aviva_EN indb 98 2 22 2018 10 08 42 AM ...
Page 116: ...110 Select Save 12 73627_07030355001_F_Aviva_EN indb 110 2 22 2018 10 08 50 AM ...
Page 126: ...120 Select Save 15 73627_07030355001_F_Aviva_EN indb 120 2 22 2018 10 08 57 AM ...
Page 132: ...126 73627_07030355001_F_Aviva_EN indb 126 2 22 2018 10 09 02 AM ...
Page 208: ...202 73627_07030355001_F_Aviva_EN indb 202 2 22 2018 10 09 52 AM ...
Page 212: ...206 73627_07030355001_F_Aviva_EN indb 206 2 22 2018 10 09 54 AM ...
Page 258: ...252 73627_07030355001_F_Aviva_EN indb 252 2 22 2018 10 10 14 AM ...
Page 270: ...264 73627_07030355001_F_Aviva_EN indb 264 2 22 2018 10 10 20 AM ...
Page 326: ...320 73627_07030355001_F_Aviva_EN indb 320 2 22 2018 10 10 32 AM ...
Page 350: ...344 73627_07030355001_F_Aviva_EN indb 344 2 22 2018 10 10 33 AM ...