13
273
Information Messages
Screen Message
Possible Causes
Possible Solutions
Battery low
is displayed while the meter is
plugged into a computer with a USB cable.
The computer USB port cannot charge the
meter battery.
Charge the meter battery using a power
outlet. See the
Charging the Battery
Using a Power Outlet
section in the
Startup
chapter in this User’s Manual.
Bolus above allowed amount
NOTE: This message can only occur if
bolus advice is turned on.
The recommended total bolus amount is
greater than the maximum allowed total
bolus amount. The meter sets the total
bolus amount to equal the maximum bolus
allowed. The carbohydrate bolus and the
correction bolus are adjusted to equal the
total bolus amount.
Check the accuracy of all entries. If
necessary, contact your healthcare
professional.
73627_07030355001_F_Aviva_EN.indb 273
2/22/2018 10:10:21 AM
Summary of Contents for Accu-Chek Aviva Insight
Page 2: ...73627_07030355001_F_Aviva_EN indb 2 2 22 2018 10 08 01 AM ...
Page 52: ...46 73627_07030355001_F_Aviva_EN indb 46 2 22 2018 10 08 20 AM ...
Page 78: ...72 73627_07030355001_F_Aviva_EN indb 72 2 22 2018 10 08 31 AM ...
Page 104: ...98 Select Save 9 73627_07030355001_F_Aviva_EN indb 98 2 22 2018 10 08 42 AM ...
Page 116: ...110 Select Save 12 73627_07030355001_F_Aviva_EN indb 110 2 22 2018 10 08 50 AM ...
Page 126: ...120 Select Save 15 73627_07030355001_F_Aviva_EN indb 120 2 22 2018 10 08 57 AM ...
Page 132: ...126 73627_07030355001_F_Aviva_EN indb 126 2 22 2018 10 09 02 AM ...
Page 208: ...202 73627_07030355001_F_Aviva_EN indb 202 2 22 2018 10 09 52 AM ...
Page 212: ...206 73627_07030355001_F_Aviva_EN indb 206 2 22 2018 10 09 54 AM ...
Page 258: ...252 73627_07030355001_F_Aviva_EN indb 252 2 22 2018 10 10 14 AM ...
Page 270: ...264 73627_07030355001_F_Aviva_EN indb 264 2 22 2018 10 10 20 AM ...
Page 326: ...320 73627_07030355001_F_Aviva_EN indb 320 2 22 2018 10 10 32 AM ...
Page 350: ...344 73627_07030355001_F_Aviva_EN indb 344 2 22 2018 10 10 33 AM ...