USE 09 MEASURED VALUES AND PARAMETERS
...............................................................................................................................................................................................................
...............................................................................................................................................................................................................
Manual USZ 08 · EN05 · 2015-06
40
Pressure (option)
Value
Unit
Modbus
Description
A-01 A pressure
bar_a 6252 F Indication
of
the pressure at measurement conditions
A-03 A current input
mA
6254
F Indication of the input value in mA
A-05 E p min value
bar_a
1392
F Pressure at measurement conditions, min. value
A-06 E p max value
bar_a
1394
F Pressure at measurement conditions, max. value
A-09 E p set value
bar_a
1396
F Pressure at measurement conditions, set value
A-11 E p at base cond.
bar_a
1398
F Pressure at base conditions
A-12 E curr. inp. gradient
1400
F Gradient (correction of the mA value)
A-13 E curr. inp. offset
1402
F Offset (correction of the mA value)
A-14 E p err. min
bar_a
1404
F Pressure at measuring conditions, lower error limit
A-15 E p err. max
bar_a
1406
F Pressure at measuring conditions, upper error limit
A-17 E p mode
4078
M Pressure at measurement conditions, operating mode (
OFF
, SET
VALUE, 4-20mA, 4-20mA_ERR)
Note:
If a pressure transmitter is connected to the USE 09 ultrasonic electronic system, the
measured value displayed here is not identical to that of the associated volume
corrector!
Temperature (option)
Value
Unit
Modbus
Description
B-01 A temperature
°C
6256
F PT100 input, indication of the temperature
B-03 A PT100 resistance
Ohm
6258
F PT100 input, indication in ohm
B-09 E T set value
°C
1408
F PT100 input, set value for temperature
B-11 E T at base cond.
K
1410
F Temperature at base conditions
B-12 E T gradient
1412
F Gradient (correction of the ohm value)
B-13 E T offset
1414
F Offset (correction of the ohm value)
B-14 E T err. min
°C
1416
F PT100 input, temperature, lower error limit
B-15 E T err. max
°C
1418
F PT100 input, temperature, upper error limit
B-17 E T mode
4079
M PT100 input, operating mode (
OFF
, SET VALUE, PT100, PT100_ERR)
Note:
If a temperature transmitter is connected to the USE 09 ultrasonic electronic system, the
measured value displayed here is not identical to that of the associated volume
corrector!
USE09-C measured values
Value
Unit
Modbus
Description
C-01 A vw
&v
6220
F Indication
of
Vw
C-02 A vwc
&v
6222
F Indication
of
Vwc
C-03 A Qm
&Q
6224
F Intermediate result for Qm (with preceding sign)
C-04 A Qmb
&Q
6238
F Intermediate result for Qmb (with preceding sign)
C-05 A Qmc
&Q
6226
F Intermediate result for Qmc (with preceding sign)