AirLink 6940XP Operator’s Manual
Sensor Settings and Calibration
•
34
Turning the Sensor On and Off
This sensor setting allows each sensor to be switched on or off and should only need to be
configured in the field if a Return to Factory Defaults occurs.
However, a sensor can be switched off if its gas type is not needed for a particular location or a
replacement sensor element is not yet available.
The AirLink 6940XP is shipped with the sensors configured as below:
• 2 Gas AirLink 6940XP – Sensors 1 and 2 turned ON, Sensors 3 and 4 turned OFF
• 3 Gas AirLink 6940XP – Sensors 1, 2, and 3 turned ON, Sensor 4 turned OFF
• 4 Gas AirLink 6940XP – Sensors 1, 2, 3, and 4 turned ON, No sensors turned OFF
1. If necessary, enter the Product Settings and Configuration menu by pressing and holding
MENU
for approximately 6 seconds.
Press and release
ADD
or
SUB
to navigate to the desired sensor settings menu, then press
MENU
to select the menu.
2. Press and release
ADD
until the “Turn On/Off” menu item appears.
3. Press
MENU
to select the “Turn On/Off” menu item.
4. Press
ADD
to turn on a sensor that was set to “Off”, and press
SUB
to turn off a sensor that
was set to “On”.
5. Press
MENU
or
BACK
to return to the Sensor Settings menu.