31
•
Global Settings and Configuration
AirLink 6940XP Operator’s Manual
Return to Factory Default Settings
Returning the AirLink 6940XP to its factory default settings will reset all customization of the
device, including the zero and calibration settings of the sensor element. A factory default does
not change the gas type.
1. If necessary, enter the Product Settings and Configuration menu by pressing and holding
MENU
for 6 seconds.
Press
MENU
to select the Global Settings menu from the Product Settings and
Configuration menu.
2. Press and release
ADD
until the Return to Factory Default option appears.
3. Press
MENU
to view the Return to Factory Default option screen.
4. Press
ADD
to proceed to the Return to Factory Default confirmation screen.
If you do not wish to return the device to its factory default settings, press
SUB
to return to
the Global Settings menu. Pressing
MENU
or
BACK
will also return the instrument to the
Global Settings menu.
AirLink 6940XP Product and Configuration Factory Default Settings
Configuration
Setting
Network I.D.
5
Zero/Cal Timer
All Cleared
Sensor
Background Setting
4
Sensor Calibration Method (non-LEL versions only)
Manual
Sensor On/Off
All On
Sensor Zero
*Cleared*
Sensor Cal
*Cleared*
Sensor Radio Address
1, 2, 3, 4