Rish Insu 5Dx
28
9
TECHNICAL SPECIFICATIONS
Accuracy is indicated as [% rdg + (dgt number * resolution)] at 23 ± 5°C,< 80%HR
Vdc Voltage
Range
Resolution
Accuracy
Protection against overloads
10 - 600V
1V
(2.0
rdg+2 dgt)
CAT IV 600V to earth
Vac TRMS Voltage
Range
Resolution
Accuracy
Protection against overloads
10 - 600V
1V
(2.0
rdg+2dgt)
CAT IV 600V to earth
Insulation resistance – Intrinsic Accuracy (at the reference conditions)
Range
Test voltage
Resolution
Accuracy (*)
0.01M
- 0.19M
(**)
100V
1%rdg
(5
rdg.+7dgt)
0.20M
- 199G
100V
(5
rdg+3dgt) if
nA
Voltage
Test
Rmis
5
(20
rdg+3dgt) if
nA
Voltage
Test
Rmis
5
0.20M
- 499G
250V
0.20M
- 999G
500V
0.20M
- 1.99T
1000V
0.20M
- 4.99T
2500V
0.20M
- 9.99T
5000V
(*) Load Capacitive < 1nF
(**) Outside range indicated by IEC/EN61557
Resistance measurement – Operating Accuracy (compliance with conditions of IEC/EN61557-1)
Range
Test voltage
Resolution
Accuracy (*)
0.01M
- 0.19M
(**)
100V
1% rdg
(12
rdg+7dgt)
0.20M
- 199G
100V
(12
rdg+3dgt) if
nA
Voltage
Test
Rmis
5
(20
rdg+3dgt) if
nA
Voltage
Test
Rmis
5
0.20M
- 499G
250V
0.20M
- 999G
500V
0.20M
- 1.99T
1000V
0.20M
- 4.99T
2500V
0.20M
- 9.99T
5000V
(*) Load Capacitive < 1nF
(**) Outside range indicated by IEC / EN61557
Measurement of generated voltage (*)
Test mode
Nominal test voltages
Accuracy
FIX
100V, 250V, 500V, 1kV, 2.5kV, 5kV
-0%, +10% + 15V
ADJUSTABLE
100 - 1000V, step 25V
1000 - 5000V, step 50V
RAMP
100 - 1000V, step 25V
1000 - 5000V, step 50V
(*) Test voltage compliance with IEC /EN 61557-2
Test current
Test voltage
Test current
100 - 5000V
1mA
Test current
3mA (*)
(*) Automatically controlled test current.
Test time
Setting range
Resolution
5s – 99min 59s
1s