3 How to Connect Wire
3-1. Description of terminal plate
- 19 -
M4 is used as the terminal screws for the terminal plate. Attach an insulated ring terminal with insulating
coating for M4 to the tip of a cable for wiring.
NOTE
If you are considering the use of the RS-485 communication function, contact RIKEN KEIKI.
NOTE
Contact output 3 (SSR)
•
Contact output 3 of the calorimeter makes SSR output.
•
AC power supply is required to drive the contact output.
Converting SSR contact into dry contact
Add an external relay.
Contact output 3 of the calorimeter makes SSR output. (Load specification: 75 - 264 VAC, 20 mA - 1 A)
It is not possible for a DC input unit to obtain SSR output.
In this case, add an external AC relay to convert SSR contact output into dry contact output.
<Wiring Diagram>
•
Install an AC relay in the non-hazardous area.
•
Use a coil with specified voltage of 75 - 264 VAC and current of 20 mA - 1 A for the relay.
E.g.) MK3ZP (OMRON): Rated current 23.1 mA (100 VAC, 60 Hz)
OHC-800
(5) (6)
Relay
~
~
100 VAC
Input unit
Hazardous location
Non-hazardous location
Summary of Contents for RS-400 Series
Page 1: ...PT3E 0512 Explosion Proof Calorimeter OHC 800 Sampling Device RS 400 Series Operating Manual...
Page 6: ......
Page 7: ......
Page 8: ......
Page 10: ......
Page 11: ......
Page 12: ......
Page 14: ......
Page 15: ......
Page 16: ......
Page 19: ......
Page 20: ......
Page 23: ......
Page 24: ......
Page 25: ......
Page 28: ......
Page 30: ......
Page 33: ......
Page 34: ......
Page 35: ......
Page 36: ......
Page 37: ......
Page 38: ......
Page 40: ......
Page 41: ......
Page 42: ......