3 How to Connect Wire

 

3-1. Description of terminal plate 

 

 

- 19 - 

 

M4 is used as the terminal screws for the terminal plate. Attach an insulated ring terminal with insulating 

coating for M4 to the tip of a cable for wiring. 

 

 

NOTE 

If you are considering the use of the RS-485 communication function, contact RIKEN KEIKI. 

 

 

 

NOTE 

Contact output 3 (SSR) 

 

Contact output 3 of the calorimeter makes SSR output. 

 

AC power supply is required to drive the contact output. 

 

Converting SSR contact into dry contact 

Add an external relay. 

 

Contact output 3 of the calorimeter makes SSR output. (Load specification: 75 - 264 VAC, 20 mA - 1 A) 

It is not possible for a DC input unit to obtain SSR output. 

In this case, add an external AC relay to convert SSR contact output into dry contact output. 

 

<Wiring Diagram> 

 

 

 

 

 

Install an AC relay in the non-hazardous area. 

 

Use a coil with specified voltage of 75 - 264 VAC and current of 20 mA - 1 A for the relay. 

E.g.) MK3ZP (OMRON): Rated current 23.1 mA (100 VAC, 60 Hz) 

 

 

 

OHC-800 

(5) (6) 

Relay 

100 VAC 

Input unit 

Hazardous location 

Non-hazardous location 

Summary of Contents for RS-400 Series

Page 1: ...PT3E 0512 Explosion Proof Calorimeter OHC 800 Sampling Device RS 400 Series Operating Manual...

Page 2: ...me users but also the user who have already used the calorimeter must read and understand the operating manual before using this product This manual mainly provides descriptions on the sampling device...

Page 3: ...o Tube 27 4 1 Recommended external tubing system 27 4 2 Precautions on tubing work 29 5 How to Adjust Use Calorimeter 30 5 1 How to start the gas detector 30 5 2 How to use three way switching valve 3...

Page 4: ...bing required for OHC 800 has been arranged inside the main unit prior to shipment Sampling Device Classification Pressure gauge unit 1 MPa 2 MPa PSI double units 1 MPa is selected for the use in Japa...

Page 5: ...meter with needle valve Adjusts and indicates the bypass flow rate of the gas discharged from BYPASS OUT Display scale 0 5 5 L min 1 10 L min 2 20 L min Varies by the specification UNIT C PR Pressure...

Page 6: ......

Page 7: ......

Page 8: ......

Page 9: ...w 1 3 Outline drawing and tubing diagram for each specification 7 Common between RS 400 100 RS 400 10 and RS 400 11 Outdoor box SUS with shading plate Swagelok fitting for 1 4 tube is used at all conn...

Page 10: ......

Page 11: ......

Page 12: ......

Page 13: ...view 1 3 Outline drawing and tubing diagram for each specification 11 Common between RS 400 200 RS 400 20 and RS 400 21 Indoor box SPCC with window Swagelok fitting for 1 4 tube is used at all connect...

Page 14: ......

Page 15: ......

Page 16: ......

Page 17: ...rimeter is exposed to oil chemicals etc 2 Place with vibrations 3 Place where radio wave or noise is generated 4 Place where the calorimeter may drop or receive strong shock 5 Place exceeding the oper...

Page 18: ...nstallation Otherwise the device may be damaged CAUTION If the calorimeter is to be fixed on the wall install it properly where can hold its weight CAUTION When installing the calorimeter on a freesta...

Page 19: ......

Page 20: ......

Page 21: ...to drive the contact output Converting SSR contact into dry contact Add an external relay Contact output 3 of the calorimeter makes SSR output Load specification 75 264 VAC 20 mA 1 A It is not possibl...

Page 22: ...S 2 sq 2 core 11 0 4 20 mA line CVVS 1 25 sq 2 core 10 0 CVVS 2 sq 2 core 11 0 Contact x 1 lines CVVS 1 25 sq 2 core 10 0 CVVS 2 sq 2 core 11 0 Contact x 2 lines CVVS 1 25 sq 4 core 11 0 CVVS 2 sq 4 c...

Page 23: ......

Page 24: ......

Page 25: ......

Page 26: ...pecification Allowed time of momentary blackout Up to 40 milliseconds The calorimeter recovers from momentary blackouts that last 40 milliseconds or longer by restarting itself Example of actions Inst...

Page 27: ...caused by lightning is impossible the following protective measures can be taken Protection against lightning Take appropriate measures in accordance with the importance of the facilities and the env...

Page 28: ......

Page 29: ...w pressure line of about 20 kPa BYPASS OUT can be connected If there is no low pressure line release the gas from BYPASS OUT into the atmosphere Since there is no limit on the arrival time of the refe...

Page 30: ......

Page 31: ...essure 0 1 MPa Bypass flow rate 5 10 L min Setting pressure 0 1 MPa Bypass flow rate 10 L min 10 3 8 Setting pressure 0 1 MPa Bypass flow rate 10 20 L min Setting pressure 0 1 MPa Bypass flow rate 20...

Page 32: ...e AC or DC specifications 3 Check that the external tubing has been connected properly 4 Check that no leak occurs from a tube between the measurement location and gas inlet 5 Turn on the measuring pa...

Page 33: ......

Page 34: ......

Page 35: ......

Page 36: ......

Page 37: ......

Page 38: ......

Page 39: ...part OHC 800 from being damaged by excessive pressure The relief valve is released when the specified pressure 10 kPa or more is applied to the valve and the pressure is transferred to the RELEFE OUT...

Page 40: ......

Page 41: ......

Page 42: ......

Page 43: ...limits Adjust the flow rates using the respective needle valve 6 2 Procedures to store the calorimeter or leave it for a long time 1 If the calorimeter is to be left unused for a long time expel the...

Page 44: ...alorific value density and WOBBE index on this product Reference gas Gas used on an optical sensor unit as a reference for measuring the refractive index of a measuring gas The gas type varies dependi...

Page 45: ...on with Contact Output 1 Red Lights up in conjunction with Contact Output 2 Green Lights up in conjunction with Contact Output 3 External Output 1 4 20 mA DC insulated source current type maximum load...

Page 46: ...ss Measuring gas temperature Same as ambient temperature Outer dimensions Approx 286 W x 453 H x 150 D mm Weight Approx 23 kg Explosion proof structure Flame proof enclosures Explosion proof class Ex...

Reviews: