3 How to Connect Wire
3-2. Recommended cable
- 20 -
3-2. Recommended cable
Connected to
Recommended cable
Cable overall outer diameter
Power (AC) line
CVV 1.25 sq/3-core
Ø10.0
CVV 2 sq/3-core
Ø11.0
Power (DC) line
CVVS 1.25 sq/2-core
Ø10.0
CVVS 2 sq/2-core
Ø11.0
4 - 20 mA line
CVVS 1.25 sq/2-core
Ø10.0
CVVS 2 sq/2-core
Ø11.0
Contact x 1 lines
CVVS 1.25 sq/2-core
Ø10.0
CVVS 2 sq/2-core
Ø11.0
Contact x 2 lines
CVVS 1.25 sq/4-core
Ø11.0
CVVS 2 sq/4-core
Ø12.0
Contact x 3 lines
CVVS 1.25 sq/6-core
Ø13.0
CVVS 2 sq/6-core
Ø14.0
RS485 line
Shielded cable of KPEVS etc.
0.75 sq/2P
Ø11.0
NOTE
The overall outer diameters must be checked because they may slightly vary between manufacturers.