1.PRODUCT OUTLINE

 

 
1-1.

Preface

 

Thank you for purchasing our fixed type Oxygen Detector Head GD-F3A-SC-A. 
First  of  all,  verify  that  the  model  number  of  this  product  is  identical  with  specifications 
described in this manual. This manual is a guide to operation of the GD-F3A-SC-A.    All 
persons who use unit for the first time and who has ever used this unit are requested to read 
through the manual to understand the content before use. 

 
1-2.

Application for use

 

This is a diffusion type oxygen detector head applying the galvanic cell type oxygen sensor 
and is used in combination with indicator / alarm unit to show oxygen content.    Read the 
instruction manual for indicator / alarm unit as well as this manual.   
This unit is designed as drip-proof construction for outdoor installation.   
Also, it can be used as intrinsically safe system in combination with zener barrier (option) to 
be able to install in hazardous area. 

 

1-3.

Identification of each signal word 

This manual contains the following headings to ensure the safe and effective operation. 

   

Indicates  an  imminently  hazardous  situation  which,  if  not  avoided, 
will  result  in  death  or  serious  injury  or  serious  damage  to  the 
product. 
This signal word is to be limited to the most extreme situation. 

 

                                                  Indicates a potentially hazardous situation which, if not avoided, 

could result serious injury to the human body or object. 

 

 

                                                          Indicates a potentially hazardous situation which, if not avoided, may   

result in minor or moderate injury or some damage to the human body 
or objects. It may also be used to alert against unsafe practices.   

 
 
                                                This means  “ADVICE”  regarding operation. 
                                                         

 

      DANGER 

      WARNING 

      CAUTION 

  *    NOTE 

Summary of Contents for GD-F3A-SC-A

Page 1: ...Oxygen Gas Detector Head GD F3A SC A Operating Manual PT2 150 PT2E 2490...

Page 2: ...tor head 2 Alarm level for gas alarm instrument shall be set within the performance of the instrument Our standard alarm level is 18 by volume for the prevention of oxygen deficiency 3 This is not con...

Page 3: ...le 10 Example of cable connection 10 OPERATION METHOD Preparation before start up 12 Start up 12 Stopping the operation 12 MAINTENANCE AND INSPECTION Inspection frequency and items 13 Calibration Requ...

Page 4: ...L CASE Responding to trouble alarm 18 Before it is thought to be a trouble 18 PRODUCT SPECIFICATIONS Standard specifications 19 Standard accessories 19 Optional accessories 19 Detection principle 20 D...

Page 5: ...esigned as drip proof construction for outdoor installation Also it can be used as intrinsically safe system in combination with zener barrier option to be able to install in hazardous area Identifica...

Page 6: ...nterval Do not change or remodel the circuit structure of this unit Do not disassemble the sensor since there is a electrolyte inside If your finger touches the electrolyte wash your finger away with...

Page 7: ......

Page 8: ...11mm from outside Hinge Hinge for front door Lock Lock for front door Mounting hole Mounting hole to fix this unit on the wall surface Terminal strip 2 Relayed terminal for input output signal use to...

Page 9: ...apacitance Co of intrinsic safety circuit and external wiring inductance Lw and capacitance Cw of intrinsic safety circuit Lo Li Lw Co C Cw 3 When making wiring connection for intrinsic safety circuit...

Page 10: ......

Page 11: ......

Page 12: ......

Page 13: ...cable For system installation operation and maintenance take care not to damage the intrinsically safe design The power of the indicator alarm unit shall be turned on after completion of all wirings...

Page 14: ...ted correctly this system starts operation by putting on the power switch in indicator alarm unit Be sure to perform gas calibration at the time of new installation Correct reading may not be obtained...

Page 15: ...cessary to carry out gas calibration at least every 6 months The law in Japan sets forth that the unit must issue an alarm during the circuit inspection related to alarm while providing normal operati...

Page 16: ...the supply of zero gas disconnect calibration adapter and confirm that the oxygen reading returns to 20 9vol Completion of zero adjustment Performing a span adjustment Plug the ammeter lead into termi...

Page 17: ......

Page 18: ...sor Environment of low temperature 5 10 Environment of low oxygen content Put a sensor into the container and seal off the container List of recommendable spare parts for regular replacement No Name o...

Page 19: ...etector head is relocated select a new place in accordance with 4 2 Installation place and 4 4 Installation works For information on wiring work see 4 5 Caution at wiring construction The unpowered ti...

Page 20: ...agent or RIKEN KEIKI Before it is thought to be a trouble Check the wirings around the related instruments as well as this detector head It may be caused by improper operation Re check the detection...

Page 21: ...m unit Zener barrier MTL7728ac MTL7728 MTL7728 ZB 1A ZB 1A ZB 1A ZB 4A ZB 4A ZB 4A Cable entry Cable gland Applicable cable O D 8 11mm Operating temperature 10 40 non rapidly vary Operating humidity L...

Page 22: ......

Page 23: ...r Safety retainer for explosion protection by limiting the energy of electric current within a non ignition level CVVS This is a kind of connecting cable between indicator alarm unit and oxygen detect...

Page 24: ...Manual Log Rev Amendment Issue data 0 First issue 2018 1 9...

Reviews: