3-3.System composition 

 

(1) Intrinsically safe system 
 
 
 
 
 
 
 
 
 
 
 

 

 
 
 

1. The oxygen detector shall be used as following composition. 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

2. The safety retainer to be connected with oxygen detector shall be passed the type approval as it is. 

        Also it shall satisfy the following conditions. 

(1)

 

Rating of safety retainer 

      Max voltage of intrinsic-safety circuit: Below 28V 

      Max current of intrinsic-safety circuit: Below 93 mA 

      Max power of intrinsic- safety circuit: Below 0.65W 

 

(2)

 

Performance division and group 

      Performance division: ia 

      Group :  ⅡC 

 

(3)

 

Relations between allowable inductance (Lo) and allowable capacitance (Co)   

      of intrinsic-safety circuit, and external wiring inductance (Lw) and capacitance(Cw)   

      of intrinsic-safety circuit. 

              Lo  ≧  Li + Lw 

              Co  ≧  C + Cw 

 

3.  When  making  wiring  connection  for  intrinsic-safety  circuit,  take  care  not  to  induce  an  electric 

current  and  voltage  which  will  spoil  the  performance  of  intrinsically  safe  by  the  electromagnetic 

induction or electrostatic induction to the intrinsic-safety circuit.

 

WARNING 

Detector 
head 

Zener barrier 

Indicator

/alarm 

unit 

Hazardous area 

Non-hazardous area 

In-between head and zener barrier:   

CVVS 1.25mm

2   

2-core or equivalent(max 250m) 

However, that of head and indicator/alarm unit: 

CVVS 1.25mm

2   

2-core or equivalent(max 2km) 

4-20mA 

 

Hazardous  
area 

Non-hazardous 
area 

Safety retainer 

Oxygen detector 
Li≒ 0 
Ci≒ 0 
≒ 0 shows a negligible value

Li 
Ci 

Lo
Co

Lw
Cw

Summary of Contents for GD-F3A-SC-A

Page 1: ...Oxygen Gas Detector Head GD F3A SC A Operating Manual PT2 150 PT2E 2490...

Page 2: ...tor head 2 Alarm level for gas alarm instrument shall be set within the performance of the instrument Our standard alarm level is 18 by volume for the prevention of oxygen deficiency 3 This is not con...

Page 3: ...le 10 Example of cable connection 10 OPERATION METHOD Preparation before start up 12 Start up 12 Stopping the operation 12 MAINTENANCE AND INSPECTION Inspection frequency and items 13 Calibration Requ...

Page 4: ...L CASE Responding to trouble alarm 18 Before it is thought to be a trouble 18 PRODUCT SPECIFICATIONS Standard specifications 19 Standard accessories 19 Optional accessories 19 Detection principle 20 D...

Page 5: ...esigned as drip proof construction for outdoor installation Also it can be used as intrinsically safe system in combination with zener barrier option to be able to install in hazardous area Identifica...

Page 6: ...nterval Do not change or remodel the circuit structure of this unit Do not disassemble the sensor since there is a electrolyte inside If your finger touches the electrolyte wash your finger away with...

Page 7: ......

Page 8: ...11mm from outside Hinge Hinge for front door Lock Lock for front door Mounting hole Mounting hole to fix this unit on the wall surface Terminal strip 2 Relayed terminal for input output signal use to...

Page 9: ...apacitance Co of intrinsic safety circuit and external wiring inductance Lw and capacitance Cw of intrinsic safety circuit Lo Li Lw Co C Cw 3 When making wiring connection for intrinsic safety circuit...

Page 10: ......

Page 11: ......

Page 12: ......

Page 13: ...cable For system installation operation and maintenance take care not to damage the intrinsically safe design The power of the indicator alarm unit shall be turned on after completion of all wirings...

Page 14: ...ted correctly this system starts operation by putting on the power switch in indicator alarm unit Be sure to perform gas calibration at the time of new installation Correct reading may not be obtained...

Page 15: ...cessary to carry out gas calibration at least every 6 months The law in Japan sets forth that the unit must issue an alarm during the circuit inspection related to alarm while providing normal operati...

Page 16: ...the supply of zero gas disconnect calibration adapter and confirm that the oxygen reading returns to 20 9vol Completion of zero adjustment Performing a span adjustment Plug the ammeter lead into termi...

Page 17: ......

Page 18: ...sor Environment of low temperature 5 10 Environment of low oxygen content Put a sensor into the container and seal off the container List of recommendable spare parts for regular replacement No Name o...

Page 19: ...etector head is relocated select a new place in accordance with 4 2 Installation place and 4 4 Installation works For information on wiring work see 4 5 Caution at wiring construction The unpowered ti...

Page 20: ...agent or RIKEN KEIKI Before it is thought to be a trouble Check the wirings around the related instruments as well as this detector head It may be caused by improper operation Re check the detection...

Page 21: ...m unit Zener barrier MTL7728ac MTL7728 MTL7728 ZB 1A ZB 1A ZB 1A ZB 4A ZB 4A ZB 4A Cable entry Cable gland Applicable cable O D 8 11mm Operating temperature 10 40 non rapidly vary Operating humidity L...

Page 22: ......

Page 23: ...r Safety retainer for explosion protection by limiting the energy of electric current within a non ignition level CVVS This is a kind of connecting cable between indicator alarm unit and oxygen detect...

Page 24: ...Manual Log Rev Amendment Issue data 0 First issue 2018 1 9...

Reviews: