19
9.PRODUCT SPECIFICATIONS
9-1.Standard specifications
Model
GD-F3A-SC-A
Detection principle
Galvanic cell method
Detectable gas
O2
Detection method
Diffusion method
Transmission method
Analog transmission
Transmission specification
4-20mADC
Transmission cable
CVVS 1.25mm
2
・2-core or equivalent
Transmission distance
Up to 2km with CVVS 1.25mm
2
(up to 250m between the detector head and Zener barrier)
Power supply
24VDC (supplied from indicator/alarm unit)
Zener barrier
MTL7728ac//MTL7728-
ZB-1A/ZB-1A+/ZB-1A-
ZB-4A/ZB-4A+/ZB-4A-
Cable entry
Cable gland(Applicable cable O.D.φ8~φ11mm)
Operating temperature
-10~+40℃ (non-rapidly-vary)
Operating humidity
Less than 90%RH (non-condensing)
Structure
Wall mounting type/mountable to pipe size 2B<50A>
Explosion-proof structure
Intrinsically safe structure(by using zener barrier)
Explosion-proof grade
Ex ia ⅡC T4 X
Outer dimension
Approx. 140(W)×175(H)×95(D)mm
Weight
Approx.1.6kg
Color
Munsel 2.5Y9/2 half tone
9-2.Standard accessories
・Operating instruction manual
9-3.Optional accessories
・U-bolt (for nominal diameter 2B)
・Calibration adopter
・Zener barrier (MTL7728ac)