Chapter 2 Functions of the Front Panel
RIGOL
VNA User Guide
2-29
Users can set the customized calibration kits. The parameters for different standard
kits are shown in the following table.
Table 2-25 Parameters for Different Standard Kits
Parts Type
Offset
Length
R
C
Parameter
L
Parameter
Att
Open
0.0000mm
-
0.00
-
-
Short
0.0000mm
-
-
0.00
-
Match
-
[1]
50.0
-
-
-
Through
0.0000mm
-
-
-
0.00
Note
[1]
:
- indicates that the parameter for this type of the standard kits is not available.
The default value, range, unit, and step value of each parameter are shown in the
following table.
Table 2-26 Parameter Value
Parameter Offset
Length
R
C
Parameter
L
Parameter
Att
Default
0.00
50.0
0.00
0.00
0.00
Range
0.00 G to
100 G
0.00 G to
100 G
0.00 to 100
G
0.00 to 100
G
0.00 to
100 G
Unit
mm
Ω
C0: fF
C1: fF/GHz
C2: fF/GHz
2
C3: fF/GHz
3
C0: pH
C1: pH/GHz
C2: pH/GHz
2
C3: pH/GHz
3
dB
Step
at 1-2-5 step at 1-2-5
step
at 1-2-5 step at 1-2-5 step at 1-2-5
step