Chapter 2 Functions of the Front Panel
RIGOL
VNA User Guide
2-13
TG
Sets the parameter related to the tracking generator (TG).
Output
Enables or disables the tracking generator. In VNA mode, TG is enabled by default.
When performing the calibration, TG is required to be enabled. After TG is disabled,
the analyzer will stop sweeping.
When the tracking generator is enabled, a signal with the same frequency of the
current sweep signal will be output from the [
GEN OUTPUT
50Ω
]
connector on the
front panel. The power of the signal can be set through the menu.
Amplitude
Sets the output power of the TG signal.
Table 2-13 Tracking Generator Output Signal Amplitude
Parameter
Remarks
Default
S11/S21: -10 dBm
DTF: 0.00 dBm
Range
-40 dBm to 0 dBm
Unit
dBm, -dBm, V, mV, uV
Knob Step
1 dB
Left/Right Arrow Key Step
Up/Down Arrow Key Step
10 dB
Trace
Displays the sweep signal on the screen. Press
Trace
on the front panel to enter the
trace setting menu. You can also tap the function keypad to select Trace to set the
relevant parameters of trace.
Selected Trace
Selects the current trace.
In VNA mode, at most 4 traces can be displayed simultaneously. Each trace is
indicated by a different color.
⚫
Trace1: yellow
⚫
Trace2: blue
⚫
Trace3: green