RIGOL
Chapter 2 Functions of the Front Panel
2-26
VNA User Guide
Remarks:
⚫
Before calibration, set the required sweep frequency range first and select the
specified calibration kit.
⚫
The calibration is only valid for the S11 measurement item. When switching to
S21 or DTF, you need to recalibrate.
Clear
Clears the calibration data acquired for the current measurement.
Port Extension
Extends the calibrated plane by setting the delay and loss of the test signal resulting
from the electrical length and compensation fixture of the connector. In this way, you
do not need to perform extra calibration any more.
Port Extension
Enables or disables the port extension function. By default, it is disabled.
Only when the port extension function is enabled, can you set the parameters for
port extension.
Port 1 Delay
Sets the delay time for Port 1 extension.
The relationship between port delay and port length is correlative, as shown in the
following formula:
Port Length = Port Delay * Velocity of Light * Velocity Factor
Table 2-20 Port 1 Delay
Parameter
Remarks
Default
0.00 us
Range
-10.00 s to 10.00 s
Unit
s, ms, us, ns
Knob Step
Port 1 delay/100
Left/Right Arrow Key Step
Up/Down Arrow Key Step
Port 1 delay/10
Unit
Sets the length unit for port extension: meter or feet. The default unit is meter.
Port 1 Length
Sets the extension length for Port 1. You should set it based on the actual parameters
of the cable for the port.