63
4-4-5 AUTO QUICK Mode
Refractive measurements may be difficult to obtain due to rapid eye movements in patients
with nystagmus or children, or for other reasons. In such a situation, the AUTO QUICK
mode function automatically activates the quick measurement mode which requires less
time for measurement.
1.
Enter to [SETUP] with
, set <AUTO QUICK> in [OTHER] to “ON”.
2.
The case when the REF measurement cannot be executed after 6 seconds, shift to
QUICK mode automatically.
3.
When switching to QUICK mode, “AQ” is displayed on the left side of the screen.
Measurement screen when switching to QUICK mode by AUTO QUICK mode
The AUTO QUICK mode is only enable at “AUTO measurement mode (A)” and
“AUTO measurement mode 2(A2)”. It is unable when “Continuous measurement
mode (CONT)”.
4.
After entering QUICK mode by AUTO QUICK mode, it is automatically released under
the following conditions.
Condition
1
Measurement of one eye ends.
2
Measurement suspended by pressing measurement switch.
3
The right/left eye discrimination is switched.
AUTO QUICK mode