55
4.
If the pupil diameter less than 2.3mm, REF measurement value is displayed as “
--
”.
5.
When < P-S > is set to “ONE”, if the pupil size measurement value is less than 2.3 mm,
pupil size measurement is performed again, and if it is 2.3 mm or more, normal
measurement is performed. Measure the pupil size 3 times and if all diameters are
less than 2.3 mm, measurement of one eye is ended and an error display is made as
follows.
Error indication when data cannot be obtained
because the pupil size is less than 2.3 mm
6.
When < P-S > is set to “ALL”, measurement is performed 5 times. The variation is
judged from REF measurement value when the pupil size is 2.3 mm or more, and if
stable data is obtained, the measurement will be completed. Measure the pupil size 5
times and if all diameters are less than 2.3mm, measurement of one eye is finished
and the error display just as when <P-S> is set to “ONE” is made.
Error indication when data cannot be obtained
because the pupil size is less than 2.3 mm
7.
If REF measurement value cannot be obtained because the pupil size is less than
2.3mm, an error is displayed on measurement completed screen as follows.