49
48
PRECAUTION
- When you open a new bottle of
Solution, please
write the discard date
on the label.
Solution is good for
3 months
after opening the bottle, or
until the expiration date printed on the label, whichever comes first.
Control
Control
- Replace the vial cap of control solution and close tightly right away after using of control
solution sample.
- Check the expiration date before you use the control solution. ( Refer to Control Solution
Package Insert. )
Exp. Date 2017-12
About Quality Control Test
Starting with Check Key
1. Insert the Check Key with " Check Key " word, facing up, into test strip
port. The meter quickly goes through a series of internal checks.
2. The check key allows you to make sure your meter is functioning well. You can
see " " symbol and " " symbol blinking on the screen.
The check key also allows your meter proceeding quality control test with control
solution.
NOTE
The check key is only for functional testing of meter. If you want to make sure
strip and meter work properly, please use control solution for the test ( see
page 50 for detail. )
3. If both " " symbol and " " symbol blinking appear, remove check key
and insert it again. If error screen appears again, your meter is not working
correctly. Please contact Bionime authorized distributor or Customer Service
Center.
4. If you don't want to conduct the control solution test after the test of check key,
please press " " to leave " " mode first. Then start the blood test procedure.
Otherwise, your test resuit will be recorded as " " data which will not be
counted in average data.
Performing a Quality Control Test
Example
2017
-
12