8. When you hear a sound of " beep " , leave the meter on the table while waiting
the test result. The screen will show the countdown mode.
9. Tightly recap the cap of control solution.
10. The control result appears after the measurement is completed. Compare your
control test result to the range printed on strip vial label.
Your result should fall within the control solution
range printed on the label of strip vial. If result is
not within the control range, please refer to
Understanding Control Test Results
on page 54.
Please follow the local regulation and discard the
used strip properly.
Performing a Quality Control Test
Performing a Quality Control Test
PRECAUTION
- Our suggestion environment for Control Solution Test is 15 °C~40 °C(59 °F ~104°F).
- Your quality control test result will not be calculated for average reading but can be
recalled. The control solution test result will be shown with " " symbol on the screen.
- You have to insert check key test before control solution test. If the quality test is performed
without " " symbol, the test result will be memorized and calculated for average same as
real blood test.
- Don't drip the Control Solution to sample entry of the strip directly. The
reagent on strip might be sucked into the bottle of Control Solution
and might cause the degeneration of Control Solution. Doing this
might contaminate the meter via the test strip port as well.
- Don't touch the tip of the control solution bottle. If you have touched it,
please clean up with water.
- Keep your Smart Code Key dry and clean. Please clean it up right
away if your Smart Code Key gets stain or moisture. ( Refer to page
55 for cleaning )
Example
CLOSE
Control Solution Range
90-130 mg/dL
5.0-7.2mmol/L
×