Your control solution test results should fall within the control solution range. That means your
System is working correctly.
If control solution test results are out of control solution range, your System may not be
working properly. Repeat the quality control test. If your control solution results outside the range still
exist, do not use the System to test your blood glucose. And contact Bionime authorized
distributor or Bionime Customer Service.
- our control solution has expired or after 3 months since opened.
- Your test strip has expired.
- You leave the cap off the vial of test strips or control solution for a long time.
- You didn't perform the test proce
-
The reason your control solution results are out of the range :
Y
dure correctly.
Malfunction of the meter
.
Understanding Control Test Results
Caring for Your Meter
Maintenance
Keep your meter and test strip free of dust, water or any other liquid. Store the meter in the carrying
case when not in use. If meter is dropped or damaged, perform a quality control test with the check
key and control solution before doing a blood glucose test.
Cleaning Meter
Clean the outside of the meter with a damp cloth and mild soap/detergent. Keep the test strip port
and Smart Code Key base from getting wet.
Cleaning Smart Code Key
If your Smart Code Key is stained with blood, control solution or any liquid, please use a dry tissue or
alcohol swab to clean it up immediately. Do not use any thing wet to clean it. Perform a check key
and quality control test to ensure the system is working properly.
Control Soultion Range
90-130 mg/dL
5.0-7.2 mmol/L
Example of control solution range
printed on your test strip vial label.