Setting and Adjustment
42
MiniFlex 600: Benchtop X-ray Diffractometer
Optics alignment (D/teX Ultra)
This section describes how to perform the alignment of D/teX Ultra after the
optics alignment of SC is completed. If the D/teX Ultra is removed from the
system and re-installed, perform the alignment again.
D/teX Ultra is an optional
component.
1
Remove the SC and install the D/teX Ultra.
For how to install the D/teX Ultra, see “
”
(Fig. 5.12)
①
End the system by following the instructions of "Shutting down the
system". (see “
Then, remove the radiation enclosure.
②
Remove the SC cable from the SC, and then remove the SC by the
whole SC stage.
③
Loosen the fixing screws of the receiving slit box, and slide the
receiving slit box to the sample side. Then, tighten the fixing screws.
④
Replace the SC with D/teX Ultra.
⑤
Bundle the SC cable and wind it behind the D/teX Ultra to avoid
interfering with the goniometer arm.
⑥
Loosen the fixing screws of receiving slit box, and slide the receiving
slit box to the D/teX Ultra side. Then, tighten the fixing screws.
Fig. 6.2.18 Moving the receiving slit box
Receiving slit
box fixing screw
Slide to the
D/teX Ultra side