RF800 [Revision 1.0.0] 01.08.2018
Shafts Geometry Measurement Machine. RF800 Series
19
8.2.1. Shafts of different diameters
The machine is equipped with the Morse tapers of two types: a small taper and a
big taper (see Figure 5 and Table 5 in Catalog RF800.00.000, items 7 and 8). To replace
the tapers, it is necessary to: unscrew the calibers, get the tapers, tighten the calibers,
install the new tapers.
8.2.2. Shafts of different lengths
To measure the shafts of different lengths, the machine is adjusted by moving the
upper rack of the Morse taper with the end sensor fixed to it along the table (see Figure 5 in
Catalog RF800.00.000).
8.2.3. Setting the origin of the linear coordinate
After you replace the support taper, it is necessary to set the origin of the linear
coordinate.
The upper end sensor, installed on the movable Morse taper, specifies the origin of
the linear coordinate of the linear translation system (the coordinate of the starting position
of the micrometer). In the starting position, the micrometer must be positioned so that the
caliber (the part of the Morse taper) is placed within the measurement zone of the
micrometer.
Follow the steps below:
1. Start
RF701_Controller.exe
, which is in the
Riftek install \RF701_Controller
directory on the desktop of the mini-PC.
2. In the
Settings
menu, select a port:
COM4
.
3. Select
Connection > Connect
in
order to connect to the system.