On the rear of the stack (where the water
comes out), you can disconnect the
flexible hose.
Disconnect the hose.
Let the water run out.
Turn the head of the hose over.
Pour the saline solution (2M NaCl)
1
into the hose.
Turn the head of the hose back over.
Reconnect the hose.
The stack must not be allowed to dry out. The stack is kept moist during
transport and storage with a saline solution. Remove this saline solution
before use and pour it back if the stack is not in use for an extended period of
time.
1
A 2 molar sodium chloride solution (2M NaCl) should be used as the saline solution.
You can make this yourself by dissolving 100 grams of table salt per litre of lukewarm
water.