Tracer AV Installation Manual - Configuration
11
3.1.3
Front Panel Configuration
The
Tracer AV
can be configured using a series of menus viewed on the RD-1 display when the
Tracer AV
is in
setup mode. Table 3-1 summarizes the functions of each of the main menus.
The two-button numeric editor works as follows:
Press
Units
to navigate RIGHT. Press
Zero
to drop DOWN or select from a list of values.
Figure 3-2. Setup Mode Menu Navigation
3.1.4
Setting Numeric Values For Parameters
1. Press
Zero
to immediately accept the current value and navigate to the next menu item.
2. Press
Units
to enter edit mode
.
The leftmost digit will flash to indicate it can be altered.
3. Press
Zero
to increment to the desired value.
4. When the value is correct, press
Units
to move to the next digit.
Menu
Menu Function
DEFCAL
Default
Configuration
Select a 500lb or 300lb default configuration. You can also display and edit zero calibration a/d
count value, test weight value, and span calibration a/d count value.
CONFIG
Configuration
Configure grads, zero tracking, zero range, motion band, overload, tare function, sample rate,
and digital filtering parameters.
FORMAT
Format
Set format of primary and secondary units, display rate.
CALIBR
Calibration
Calibrate
Tracer AV
. See Section 4.0 on page 23 for calibration procedures.
SERIAL
Serial
Configure EDP and printer serial ports.
PROGRM
Program
Set power-up mode, regulatory mode, and consecutive number values.
SETPNT
Setpoint
Configure Setpoints and digital outputs.
VERS
Version
Display installed software version number.
Table 3-1.
Tracer AV
Menu Summary
Button
Function
Zero
Move DOWN/ Increment Value
Units
Move RIGHT/Next
Table 3-2. Button Functions in Setup Mode
1
st
Level
Parameter
Default value
Value
1
st
Level
Parameter
2
nd
Level
Parameter
2
nd
Level
Parameter
Value
Value
When moving through values below the first menu, press UNITS to view the
selections. Press ZERO to set the parameter and move to the next parameter on the
leval above.