DC-400 Counting Scale
40
Visit our website
607
A/D Board type for Scale-1
0
Normal
1
Prevent from small vibration/fast change in display
2
Prevent from medium vibration
3
Prevent from large vibration/slow change in display
608
External Scale B connection
0
Disable
1
Enable
609
Minimum display (increment) for Scale B
0
2
1
1
2
5
3
10
610
Weight decimal point position for Scale B
0
0
1
0.0
2
0.00
3
0.000
4
0.0000
611
Load cell sensitivity selection (mV/V) for Scale B
0
4.00 mV/V
1
3.76 mV/V
2
3.52 mV/V
3
3.28 mV/V
4
3.04 mV/V
5
2.80 mV/V
6
2.56 mV/V
7
2.32 mV/V
8
2.08 mV/V
9
1.84 mV/V
10
1.60 mV/V
11
1.36 mV/V
12
1.12 mV/V
13
0.88 mV/V
14
0.64 mV/V
15
0.40 mV/V
612
Load cell type for Scale B
0
Standard/Normal load cell
1
Abnormal load cell with large offset
613
A/D Board type for Scale B
0
Normal
1
Prevent from small vibration/fast change in display
2
Prevent from medium vibration
3
Prevent from large vibration/slow change in display
614
External Scale C connection
0
Disable
1
Enable
SPEC Option
Description
Table 6-2. Weight and Measurement Specification Settings (Continued)