34
BenchPro Series Operation Manual
6.0
Maintenance and Troubleshooting
Prior to calling customer support have the software type and software version number available. These are
displayed briefly when powering on the scale.
6.1
Troubleshooting
Table 6-1. Error Codes
Error Code
Description
Possible Cause
Corrective Action
Calibration error
Incorrect calibration weight value used;
damaged load cell
Repeat calibration
Replace load cell
Power up or initial zero error
Upon power up, weight or item on the
plat form is greater than
setting
Remove weight and power cycle
the scale
Semi-auto zero error
When pressing the zero button, the
weight value displayed is greater than the
% in Zero configuration
--
Configuration error
Invalid configuration settings
Check configuration settings
Overload error
Too much weigh applied
Perform calibration, check LC mV
Memory error
PCB is corrupt
Replace main PCB
Low battery
Battery power voltage is below 4.2 V
Replace batteries
A/D Conversion error
--
Calibrate, replace main PCB
Diagnostics Menu
The diagnostic menu (
) is used to troubleshoot scale operation. Use DIV-A or DIV-O to test the functionality
of the load cell.
1. From the diagnostics menu, press DIVA or DIVO.A value is displayed.
2. Add weight onto the scale to see if the counts increase.
3. Remove the weight to see if the value returns the value displayed in
4. Calibrate the scale before determining a load cell is bad.
6.2
Load Cell Wiring
E+ Gr
een
E- Black
S+ Red
S- White
GND Y
ellow
Figure 6-1. Load Cell Wiring