14
BenchPro Series Operation Manual
4.0
Calibration
4.1
Span Calibration
The
BenchPro Series
allows for calibration with weight values other than full capacity.
alternate calibration weights for each model.
Calibration should only be performed using certified tests weights and performed by the local scale distributor.
Note
Turn off Gravity Compensation (
) prior to performing a calibration using certified
weights. The default capacity (
) setting is in lb, if using kg calibration weights, change the
to the
appropriate scale capacity of the scale model. The model number of the scale is located on the serial tag on
the bottom of the scale.
1. To enter the calibration mode, remove the weight platter and the main pcb cover plate.
2. Locate and press the service setup switch.
3. Place the weight platter back onto the scale.
is displayed
4. Press
until
is displayed.
5. Press
to accept.
or
is displayed.
6. Press
to scroll to
7. Press
to accept.
is displayed.
8. Press
.
is displayed.
9. Use
to choose the appropriate scale capacity.
10. Press
to scroll to chosen capacity.
11. Press
to accept.
is displayed
12. Press
until
is displayed.
13. Press
to accept.
is displayed.
14. With no weight on the weight platter, press
to accept. A six digit value displays. This is the internal
counts of the load cell at zero weight.
15. Press
to accept and perform a zero calibration.
is displayed.
16. Press
once.
is displayed.
Model
Scale Capacity
Alternate Calibration
Weights
Scale Capacity
Alternate Calibration
Weights
BP 1214-6R
15 LB
5, 10, 15LB
6 KG
2, 5, 6 KG
BP 1214-15R
30 LB
10, 20, 30 LB
15 KG
5, 10, 15 KG
BP 1214-35P
70 LB
20, 50 70 LB
35 KG
10, 20, 30 KG
100 LB
30, 50, 100 LB
50 kg
10, 25, 50 KG
BP 1214-75S
150 LB
50, 100, 150 LB
75 kg
20, 50, 75 KG
300 LB
100, 200, 300 LB
150 kg
50, 100, 150 KG
Table 4-1. Alternate Calibration Weights