380 Synergy Digital Weight Indicator
28
Visit our website
4.3.6 Setup – Print Format Menu
Figure 4-11. Setup – Print Format Menu
keyhld
Key Hold – Allows setting of the key hold time and interval
HLDTME – Key hold time (in tenths of a second); 20 equals 2 seconds;
Enter value: 10–50,
20
(default)
INTRVL – Key hold time interval; the amount of time between increments during a key hold (in twentieths of a second);
2 equals a tenth of a second (10 increments per second during a key hold);
Enter value: 1–100,
2
(default)
kybdlk
Keyboard Lock – Disables the keyboard except power;
settings:
OFF
(Default), ON
zeronl
Zero Only – Disables the keyboard except Zero and Power;
settings:
OFF
(Default), ON
bright
Brightness of Back Light; settings:
LOW
(default), MED, HIGH, OFF
Batery
Battery – Power supply via battery
NO – No battery
ALK
(default) – Alkaline battery (non-rechargable)
pwrsav
Power Saving mode – Energy saving for battery operation
NONE
(default)
LOW – Turns off backlight after 30 seconds of no activity; changing weight or a button press will illuminate the backlight
HIGH – Turns off backlight and load cell after 30 seconds at zero weight; any button must be pressed to wake up indicator
autoff
Auto Off Time – Time (in minutes) at zero weight after which the scale will power off. A setting of 0 will not power off;
Settings:
0
(default) - 60
locale
Location Gravity Compensation – Enables gravity compensation;
Settings:
OFF
(default) – gravity compensation disabled
FACTOR – uses origin and destination gravity factors to find gravity compensation
grvorg
Gravity of Origin – Original gravity factor (in m/s
2
) for gravity compensation; displays when LOCALE parameter is set to FACTOR;
Enter value: 9.00000–9.99999,
9.80665
(default)
grvdst
Gravity of Destination – Destination gravity factor (in m/s
2
) for gravity compensation; displays when LOCALE parameter is set to
FACTOR
; Enter value: 9.00000–9.99999,
9.80665
(default)
Parameter
Description
gfmt
Gross Format – Gross demand print format string
FORMAT –
Enter characters: Alphanumeric entry up to 300 characters,
Gross<g><nl2>
(default)
DEST 1 – Destination ports;
Settings:
RS232
, USB, NONE
nfmt
Net Format – Net demand print format string
FORMAT –
Enter characters: Alphanumeric entry up to 300 characters,
Gross<g><nl>Tare<sp><t><nl>Net<sp2><n><nl2>
(default)
DEST 1 – Destination ports;
Settings:
RS232
, USB, NONE
ACMFMT
Accumulator Format – Accumulator print format string
FORMAT –
Enter characters: Alphanumeric entry up to 300 characters,
Accum <a><nl>
(default)
DEST 1 – Destination ports;
Settings:
RS232
, USB, NONE
Table 4-9. Setup – Print Format Menu Description
Parameter
Description
Table 4-8. Setup – Program Menu Description
Setup
Pformt
format
dest 1
Nfmt
format
dest 1
Gfmt
Format
dest 1
Acmfmt
format
dest 1
Nfmt