Configuration
© Rice Lake Weighing Systems
●
All Rights Reserved
23
4.3.1 Setup – Configuration Menu
Figure 4-5. Setup – Configuration Menu
Parameter
Description
capcty
Capacity – Maximum rated capacity of the scale;
Enter value: 0.000001–99999.0,
10000.0
(default)
Ztrkbn
Zero Track Band – Automatically zeros the scale when within the range specified, as long as the input is within the
zrange
and
scale standstill; Max legal value depends on local regulations; Specify the zero tracking band in ± display divisions;
Enter value:
0.0–100.0,
0.0
(default)
Zrange
Zero Range – The total amount the scale can be zeroed; Zero range represents a percentage of capacity; The default value of
1.9 represents ±1.9% around the calibrated zero point, for a total range of 3.8%; A value of 0.0 prevents zeroing; Maximum legal
value depends on local regulations;
Enter value: 0.0–100.0,
1.9
(default)
Intzro
Initial Zero Range – When the indicator is turned on and the weight value is between the ± percent range specified of Calibrated
Zero, the indicator automatically zeros off the weight;
Enter value: 0
.
0–100.0,
0.0
(default)
Motban
Motion Band – Sets the level, in display divisions, at which scale motion is detected; If motion is not detected for the time defined
by
ss
time
, the motion annunciator is not displayed; Some operations, including print, tare, and zero, require the scale to be at
standstill; Maximum legal value varies depending on local regulations; If this parameter is set to 0, the motion annunciator is
never lit and operations requiring standstill are performed regardless of scale motion; If 0 is selected,
ztrkbnd
must also be
set to 0;
Enter value: 0–100,
1
(default)
Ovrloa
Overload – Determines the point at which the display blanks and the overload error message displays (^^^^^^);
Maximum legal value varies depending on local regulations;
Settings:
FS+2%
(default), FS+1D, FS+9D, FS
Smprat
Sample Rate – Selects measurement rate, in samples per second, of the analog-to-digital converter; Lower sample rate values
provide greater signal noise immunity;
Settings: 5HZ, 6HZ, 10HZ, 12HZ, 20HZ,
25HZ (Default)
, 40HZ, 50HZ, 80HZ, 100HZ
fltchn
Filter Chain Type – Sets the filter type to be used;
Settings:
AVGONLY
(default) – Digital Rolling Average Filter (
); Uses DIGFL1-3, DFSENS and DFTHRH
RAW – No filtering
digfl1-3
Digital Filters – Sets the digital filtering rate used to reduce the effects of environmental influences from the immediate area of
the scale; Settings indicate the number of A/D conversions per update which are averaged to obtain the displayed reading; a
higher number gives a more accurate display by minimizing the effect of a few noisy readings, but slows down the response time
of the indicator;
Settings: 1, 2,
4
(default), 8, 16, 32, 64, 128, 256
Dfsens
Digital Filter Sensitivity – Specifies the number of consecutive A/D readings which fall outside the Filter Threshold before filtering
is suspended;
Settings:
2OUT
(default), 4OUT, 8OUT, 16OUT, 32OUT, 64OUT, 128OUT
Dfthrh
Digital Filter Threshold – Sets a threshold value, in display divisions; when a number of consecutive A/D readings (Digital Filter
Sensitivity) falls outside of this threshold value (when compared to the output of the filter), filtering is suspended and the A/D
value is sent straight through the filter; Filtering is not suspended if the threshold is set to NONE;
Settings:
NONE
(default), 2D,
5D, 10D, 20D, 50D, 100D, 200D, 250D
RTLTRP
RattleTrap – Enables RattleTrap filtering; Effective at eliminating vibration effects, environmental influences and mechanical
interference from nearby machinery, may increase response time over standard digital filtering;
Settings:
OFF
(default), ON
Tarefn
Tare Function – Enables or disables push-button and keyed tare;
Settings:
BOTH
(default)
– Both push-button and keyed tare are enabled
NOTARE
– No tare allowed (gross mode only)
PBTARE
– Push-button tares enabled
KEYED
– Keyed tare enabled
Sstime
Standstill Time – Specifies the length of time the scale must be out of motion, before the scale is considered to be at standstill (in
0.1 sec intervals);
Enter value: 0–600,
10
(default)
Table 4-3. Setup – Configuration Menu Descriptions
Setup
Config
Capcty
fltchn
digfl1
digfl2
digfl3
dfsens
dfthrh
Ztrkbn
Zrange
Intzro
motban
ovrloa
smprat
rtltrp
tarefn
sstime