22
160-10-7 Low-Profile Digital Physician Scale Operation
To enter the program menu:
1. Enter the setup mode. See
Section 5.4 on page 20
.
2. With
PROG
displayed, press
.
3. Using the navigation instructions in
Section 5.1 on page 18
, set all
PROG
parameters.
5.5
Default Menu
The default menu is used to return the scale to the original factory settings. This will not change the calibration.
Figure 5-6. Default Menu
5.5.1
Set to Default
1. Press .
NO
displays
.
2. Press .
YES
displays.
3. Press .
DONE
displays.
4. Press
to return to main menu.
5.6
Scale Calibration
To calibrate the scale:
1. Enter the setup mode. See
Section 5.4 on page 20
.
2. Press
to scroll to
CALIB.
3. Press
. A numeric value displays (weight to be used for calibration).
4. Only calibrate in LB.
5. Press
.
The digit on the right flashes.
• Press
or
to change the value of digit.
• Press
to move to the next digit.
• When the desired value is displayed, press
.
CLEAR
displays.
OP
Binary options:
• OP0 — Live weighing options (0=disable, 1=enable)
• OP1 — Communication protocol (0=ESC, 1=maintenance)
• OP2 — BMI menu (0=disable, 1=enable)
• OP3 — RTC power (0=disable, 1=enable)
• OP4 - Semi-Auto-Live — (0=disable, 1=enable)
• OP5 - Full calculation — (0=spatial, 1=multiply by (0=disable, 1=enable
• OP6 - Tare - (0=disable, 1=enable
• OP7 - Bat type — (0=dry batteries, 1=rechargeable batteries)
• OP8 — OIML mode - (0=disable, 1=enable
SAVE
Saves settings and returns to upper level
Parameter
Description
Table 5-3. Program Mode Parameters (Continued)
ENTER
DEF
CALIB
PROG
DONE
NO
YES
ENTER
ENTER
BMI
BMI
ENTER
ENTER
BMI
ENTER