Testing and Care
Page 22 of 30
Mother Transmitters User Guide
0510-1020-O
CAUTION
: Using an approved cleaner does
not confirm the suitability of the cleaner to
meet your infection control standard; it only
indicates that the transmitter will not be
harmed by the cleaner.
The transmitters may be wiped, sprayed or
rinsed with water or the approved cleaners
during cleaning, but must not be soaked.
Soaking transmitters may result in
equipment and/or system failure.
4. Allow the cleaner to remain on the
transmitter for the amount of time
specified by the manufacturer. When
using surface disinfectants and
decontaminant cleaners, always follow
the manufacturer’s instructions.
5. Wipe the transmitter with a clean cloth
or towel, and then allow the transmitter
to air-dry. The transmitter can also be
rinsed with water, if required to in the
manufacturer’s instructions.
6. Inspect the piercing contacts on the
transmitter. The contacts must be gold
in color and not bent or out of alignment.
If the piercing contacts are NOT gold in
color or are bent out of alignment,
discard the transmitter.
7. Close the clamps on the transmitter
8. Place the clean transmitter in a storage
container for future use.