First Commissioning
18
5
First Commissioning
WARNING
W4.0002
Danger to life through electric shock
Damaged power cable
An electric shock can cause burns, cardiac arrhythmia, respiratory arrest,
as well as cardiac arrest.
•
Never use a damaged power cable to connect the device to the
mains!
•
Check the power cable and the plug for any damage before use.
NOTICE
N11.0002
Setting up the device
Disconnecting the device from the mains
A separation of the device from the mains must be possible at any time.
•
Set up the device in such a way, that the connection for the power
cable is always easily accessible.
Before first commissioning the silencer and the manual or automatic vacuum regulation must be
installed, and an industrial vacuum cleaner must be connected.
Subsequently, the desired test sieve with the appropriate lid can be inserted. The AS 200 jet is
suitable for test sieves with an outer diameter of 200 mm (with adapter ring) and 203 mm.
NOTICE
Only one
test sieve must be inserted. A large amount of sample can significantly
increase the load on the sieve mesh fabric. Make sure not to exceed the maximum payload
(
Chapter "
").
5.1 Silencer
To reduce the suction noise, it is recommended to mount the silencer (
F
) included in the
delivery to the air inlet duct (
D
).
Summary of Contents for AS 200 jet
Page 2: ...2 Copyright Copyright by Retsch GmbH Retsch Allee 1 5 42781 Haan Germany...
Page 5: ...Notes on the Manual 5...
Page 25: ...First Commissioning 25 Fig 10 Example of a label on the sieve lid...
Page 42: ...Controlling the Device 42 Fig 21 Submenu SERVICE H7 1...
Page 61: ...61...
Page 62: ...62...
Page 63: ...63...
Page 64: ...Copyright Copyright by Retsch GmbH Retsch Allee 1 5 42781 Haan Germany...