First Commissioning
24
Place the desired test sieve (
S
) on the nozzle chamber (
A
).
NOTICE
For test sieves with
200 mm outer diameter, the adapter ring (
AR
) must be used additionally.
Place the sample material on the test sieve.
Close the test sieve with the sieve lid (
R
) suitable for the sieve height and outer diameter
(
Chapter "
").
NOTICE
Only with a suitable combination of test sieve and sieve lid, the required negative
pressure can be built up in the nozzle chamber and thus the sieving process can be started.
5.5.1 Sieve Lids
For the test sieves four different types of the sieve lid are available. These differ in their shape
and their diameter, and must be chosen to match the test sieve used.
Fig. 9:
Shape of the sieve lid for a sieve height of 50 mm (top) and 25 mm (bottom)
Pay attention to the appropriate labelling on the sieve lids:
Label
Outer diameter
Sieve height
Adapter ring
8 inch x 2 inch (50 mm)
203 mm (8")
50 mm (2")
no
8 inch x 1 inch (25 mm)
203 mm (8")
25 mm (1")
no
200 mm x 50 mm
200 mm
50 mm (2")
yes
200 mm x 25 mm
200 mm
25 mm (1")
yes
50 mm (2")
25 mm (1")
Summary of Contents for AS 200 jet
Page 2: ...2 Copyright Copyright by Retsch GmbH Retsch Allee 1 5 42781 Haan Germany...
Page 5: ...Notes on the Manual 5...
Page 25: ...First Commissioning 25 Fig 10 Example of a label on the sieve lid...
Page 42: ...Controlling the Device 42 Fig 21 Submenu SERVICE H7 1...
Page 61: ...61...
Page 62: ...62...
Page 63: ...63...
Page 64: ...Copyright Copyright by Retsch GmbH Retsch Allee 1 5 42781 Haan Germany...