First Commissioning
22
5.4 Connecting an Industrial Vacuum Cleaner
WARNING
W5.0017
Danger to life through electric shock
IEC power connection for external industrial vacuum cleaner
When the device is switched on, there is a risk of an electric shock when
touching the IEC power connection for the external industrial vacuum
cleaner.
An electric shock can cause burns, cardiac arrhythmia, respiratory arrest,
as well as cardiac arrest.
•
Switch off the device before connecting the external industrial
vacuum cleaner.
CAUTION
C7.0049
Danger of objects being ejected
Connection of compressed air instead of an industrial vacuum cleaner
If compressed air is connected to one of the two air ducts, the sieve lid
and the test sieve will be ejected.
•
The device may not be operated with compressed air!
The AS 200 jet can only be operated with a suction device such as an industrial vacuum
cleaner. A suitable industrial vacuum cleaner is available as an optional accessory from Retsch
GmbH.
CAUTION
Before commissioning the industrial vacuum cleaner, read the corresponding
separate manual.
Depending on the requirement, the industrial vacuum cleaner can either be connected to the air
outlet duct (
E
) or the manual vacuum regulation (
G
) (recommended).
Connect the suction pipe (
SR
) of the industrial vacuum cleaner to the air outlet duct or
alternatively to the manual vacuum regulation (recommended).
Fig. 7:
Connecting the industrial vacuum cleaner: to the manual vacuum regulator (left,
recommended) or the air outlet duct (right)
G
E
L
SR
Summary of Contents for AS 200 jet
Page 2: ...2 Copyright Copyright by Retsch GmbH Retsch Allee 1 5 42781 Haan Germany...
Page 5: ...Notes on the Manual 5...
Page 25: ...First Commissioning 25 Fig 10 Example of a label on the sieve lid...
Page 42: ...Controlling the Device 42 Fig 21 Submenu SERVICE H7 1...
Page 61: ...61...
Page 62: ...62...
Page 63: ...63...
Page 64: ...Copyright Copyright by Retsch GmbH Retsch Allee 1 5 42781 Haan Germany...