INTRODUCTION 

 

Thank you for purchasing a 

Respond 

O

2

™ 

Ultrasonic Oxygen Indicator

The designers of this product have years of experience in the field of 
oxygen concentrator repairs.  The unreliability and high cost of operation 
of fuel cell analyzers led to the invention of the 

Respond 

O

2

™ 

Ultrasonic 

Oxygen Indicator

.  

 

By relying on a method already used in many oxygen concentrators to  
trigger "Low O 

2

" alarms – 

ultrasound

 – 

Respond 

O

2

™ 

Oxygen 

indicator offers more cost effective product to check the output of 
oxygen concentrators. The 

Respond 

O

2

 

Ultrasonic Oxygen Indicator 

responds faster, has no fuel cell to replace and is simpler to use than a 
fuel cell analyzer.   

 

Directions for Use:

 

There are two (2) controls for the 

Respond 

O

2

 

unit: 

 
1. 

 Pressing and releasing the 

“on/off”

 button will turn the unit on or 

off.  

2. The 

“pure O

2

 calibrate”

 button when pressed and held down                   

 

while connected to a compressed gas or liquid U.S.P. oxygen  

 

(> 99.0%) source verifies that the unit is properly calibrated, thus  

 

eliminating the need to have an outside source calibrate the unit. 

 

Since the 

Respond 

O

2

device uses ultrasound technology to check 

the concentrator operation, there has to be two modes of operation: 
 
1. 

When the unit is turned on, the software enables the unit to  

  measure the output of an oxygen concentrator adjusting for the       
  argon that  passes through the sieve beds. 

 

 
2. 

When using the unit with the 

“pure O

2

 calibrate”

 button  

   

depressed the software measures oxygen not adjusting for the  

   

argon. When the 

“O

2

 calibrate”

 button is released, the unit  

   

automatically returns to the concentrator mode and will then  

   

again adjust for the argon and read incorrectly for pure oxygen. 

 

 
 

 

Ultrasonic Oxygen Indicator 

 
 

 

 

 

 

INITIAL SETUP 

 

1. Remove the 

Respond 

O

2

Ultrasonic Oxygen Indicator

 from  

  

  the package.  Make sure you have the following items: 

 

•   

Respond 

O

2

indicator unit 

 

•   9-volt battery (factory installed) – Spare battery also ncluded 
•   2 foot connecting tubing

  

   

Carrying case and strap

 

•   Operation Manual 

 

2.   The 

Respond 

O

2

Ultrasonic Oxygen Indicator

 

was calibrated  

  at the factory.  However, the proper calibration of the unit's      
  oxygen concentration measurement mode can be quickly and  
  easily verified by following the "Pure Oxygen Verification"  
  procedure on 

page 6

 of this manual.  

 

3.   If accreditation organizations or internal policies and  

   

  procedures require routine calibration, instructions on  

    performing a "Quick One Step Calibration Procedure" to  

  U.S.P. oxygen (> 99.0%) can be found on 

page 6

 of this  

  manual.  A “Full Two Step Calibration Procedure" to both  
  U.S.P. oxygen (> 99.0%) and room air can be found on 

page   

  

7

 of this manual. 

 

 

Low Battery Indicator: 

When the unit displays 

“LO”

 and 

“BAT”

 in the lower left corner,  

battery is low and should be replaced.  See 

page 8

 
 
 
 
 
 
 
 
 
 
 
          

 

 

 

 

 

 

                   

3

         

Ultrasonic Oxygen Indicator 

Summary of Contents for Respond O2

Page 1: ...e 866 333 4030 Fax 866 333 4035 sales respondo2 com www respondo2 com 803350 Rev B October 2020 Respond O2 Ultrasonic Oxygen Indicator Operation Manual Responsive Respiratory Inc 6244 Lemay Ferry Rd S...

Page 2: ...turn the unit to Responsive Respiratory Responsive Respiratory will be responsible for normal shipping charges for the return of the unit to purchaser All warranties are voided if the Respond O2 unit...

Page 3: ...ration from any source other than a conventional oxygen concentrator using molecular sieve beds This unit will not withstand fluid spillage The unit will not withstand mechanical shock or vibration NO...

Page 4: ...ies Clean battery terminals When unit is turned on display flashes PRO O2 CAL and then turns off Unit requires calibration Perform Full Calibration Procedure outlined in Operation Manual Display reads...

Page 5: ...of an oxygen concentrator adjusting for the argon that passes through the sieve beds 2 When using the unit with the pure O2 calibrate button depressed the software measures oxygen not adjusting for th...

Page 6: ...s Will the unit lose calibration when I change batteries No The unit holds calibration even when the battery is removed 12 Ultrasonic Oxygen Indicator FREQUENTLY ASKED QUESTIONS The following are some...

Page 7: ...t supply tubing to the concentrator oxygen outlet and the inlet port 5 Turn on the device See above 6 The indicator will stabilize in about 10 seconds and provide an accurate reading on the LCD displa...

Page 8: ...ircuit will compensate for variations in temperature The temperature during storage and shipping should not reach below 29 F 34 C or above 160 F 71 C If the unit has been stored at other than normal r...

Page 9: ...cator FULL TWO STEP CALIBRATION PROCEDURE The following procedure may be used to calibrate the indicator to both room air and U S P oxygen 99 0 1 Do not connect unit to oxygen source or introduce oxyg...

Page 10: ...the rear of the unit away from the case Connect the 9 volt battery to the standard 9 volt battery connection and lay the battery sideways in the battery compartment Insert battery door by sliding door...