TROUBLE SHOOTING CHECKLIST (CONTINUED)

 

 

Complaint Probable 

Cause 

Suggested 
Corrective Action 

Display shows 

"err"

 during 

calibration 

Calibration was 
performed incorrectly 
or wrong calibration 
gas was used. 

Ensure U.S.P. 
oxygen (> 99.0%) is 
being used and re-
perform procedure 
according to 
directions in the 
Operation Manual. 

Display shows 

"err"

  

during use 

Unit has made a 
reading outside 
normal ranges. 

Check calibration.  
Make sure gas 
being measured is 
from oxygen 
concentrator. 

"LO BATT"

 

appears in display 
screen when unit is 
turned on 

Battery is low. 

Change battery. 

Display shows 

"err"

 or a reading 

other than 100% ± 
2% after full 
calibration 

 

Unit may have been 
used to check a 
concentrator with a 
humidifier bottle in 
line or water has 
gotten inside of unit. 

Call Responsive 
Respiratory 
Customer Service 
Department at   
(866) 333-4030 for 
authorization to 
return unit for repair. 

 
 
 
 
 
 
 
 

 

 
 
 
 
 

                  

14

 

Ultrasonic Oxygen Indicator 

 
 

 

TROUBLE SHOOTING CHECKLIST

 

 

          
 
 
                         

 
 

 

13 

Complaint Probable 

Cause 

Suggested 
Corrective Action 

Display is 

blank

 

when unit is turned 
on. 

Bad batteries. 
Corroded battery 
terminals. 

Replace batteries. 
Clean battery terminals. 

When unit is turned 
on display flashes 

"PRO O2 @ CAL"

 

and then turns off. 

Unit requires 
calibration. 

Perform "Full  
Calibration Procedure" 
outlined in Operation 
Manual. 

Display reads 

91.8%

 

on a source of 
U.S.P. oxygen (> 
99.0%). 

Unit is operating in 
mode to check 
concentrators and is 
adjusting for 
concentrator's output 
of argon. 

With unit still connected to
source of U.S.P. oxygen 
(> 99.0%) push and 
Ucontinue  to hold downU 

“pure O

B

2

B

/calibrate”

 

button to temporarily 
remove argon correction 
from unit's software. 

Display reads 
between 

98% to 102%

 (but 

not 99.9%) on a 
source of U.S.P. 
oxygen  
(> 99.0%) with 
button held down. 

Unit requires quick 
calibration. 

With unit still connected 
to source of U.S.P. 
oxygen  (> 99.0%) push 
and hold down 

“pure 

O

B

2

B

/ calibrate”

  button 

until display reads 99.9% 
- 101 (approximately 20 
seconds). The unit is then 
recalibrated. 

Unit is out of 98% to 
102% range after 
quick calibration. 

Unit requires full 
calibration. 

Perform "Full  
Calibration Procedure" 
outlined in User Manual. 

"---"

 appears on 

display while unit is 
being calibrated. 

Incomplete 
calibration. 

Release the 

"pure 

O

2

/calibrate"

 button and 

restart calibration 
process. 

Display shows 

"-U-"

Indicates an out of 
range reading. 

Perform "Full Calibration 
Procedure" outlined in 
User Manual. 

Ultrasonic Oxygen Indicator 

Summary of Contents for Respond O2

Page 1: ...e 866 333 4030 Fax 866 333 4035 sales respondo2 com www respondo2 com 803350 Rev B October 2020 Respond O2 Ultrasonic Oxygen Indicator Operation Manual Responsive Respiratory Inc 6244 Lemay Ferry Rd S...

Page 2: ...turn the unit to Responsive Respiratory Responsive Respiratory will be responsible for normal shipping charges for the return of the unit to purchaser All warranties are voided if the Respond O2 unit...

Page 3: ...ration from any source other than a conventional oxygen concentrator using molecular sieve beds This unit will not withstand fluid spillage The unit will not withstand mechanical shock or vibration NO...

Page 4: ...ies Clean battery terminals When unit is turned on display flashes PRO O2 CAL and then turns off Unit requires calibration Perform Full Calibration Procedure outlined in Operation Manual Display reads...

Page 5: ...of an oxygen concentrator adjusting for the argon that passes through the sieve beds 2 When using the unit with the pure O2 calibrate button depressed the software measures oxygen not adjusting for th...

Page 6: ...s Will the unit lose calibration when I change batteries No The unit holds calibration even when the battery is removed 12 Ultrasonic Oxygen Indicator FREQUENTLY ASKED QUESTIONS The following are some...

Page 7: ...t supply tubing to the concentrator oxygen outlet and the inlet port 5 Turn on the device See above 6 The indicator will stabilize in about 10 seconds and provide an accurate reading on the LCD displa...

Page 8: ...ircuit will compensate for variations in temperature The temperature during storage and shipping should not reach below 29 F 34 C or above 160 F 71 C If the unit has been stored at other than normal r...

Page 9: ...cator FULL TWO STEP CALIBRATION PROCEDURE The following procedure may be used to calibrate the indicator to both room air and U S P oxygen 99 0 1 Do not connect unit to oxygen source or introduce oxyg...

Page 10: ...the rear of the unit away from the case Connect the 9 volt battery to the standard 9 volt battery connection and lay the battery sideways in the battery compartment Insert battery door by sliding door...

Reviews: