background image

 
 

 

 

INSTRUCTIONS FOR CHECKING O

2

CONCENTRATORS 

 

To turn unit on and off: 

To turn the unit on or off press and release the  

“on/off”

 button on the lower left-hand side of the unit. (EXAMPLE:  

Concentrator O 

2

 20.9) When the unit is turned on the device defaults to the mode 

in which the software enables the unit to measure the output of an oxygen 
concentrator compensating for the argon that passes through the sieve beds. The 
display will read 

"PRO"

, then 

"O

2

"

 and finally  

"Concentrator O2"

 and a number.

 

 If the 

unit is hooked up to an operating 

concentrator, the device will automatically measure the  
concentration of the oxygen being produced and show the value on the  
LCD display. The unit's default operating mode automatically compensates for the 
argon that passes through the sieve beds, it will display a false reading 
(

approximately 90%

) if connected to an oxygen source other than a concentrator 

unless the specific actions discussed on 

pages 6 - 7

 of this manual are taken. If 

the device is not hooked up to an oxygen source it will measure and display the 
concentration of oxygen in the ambient air (

usually 20.9%

). 

Reading may vary if 

sample gas remains in unit

 

To turn the 

Respond 

O

2

 off, press and release the 

“on/off” 

button. The unit will 

display 

“OFF”

 and shut down. To conserve battery life, it automatically shuts 

down if any one of the unit's functions is not utilized for approximately 5 minutes.  

 

Measuring Oxygen Concentrations:

 

Follow the steps below to measure the concentration of oxygen being  
supplied by an oxygen concentrator:  
1.  Make sure the concentrator has been running long enough (follow  

           concentrator manufacturer's guidelines) to reach full operational output. 

2.  Make sure flow rate is above 2 LPM - preferably at the concentrator's  
     highest output. 
3.  

Remove any humidifier from concentrator

4.  Connect supply tubing to the concentrator oxygen outlet and the inlet   
     port  
5.  Turn on the device (See above).  
6.  The indicator will stabilize in about 10 seconds and provide an  
     accurate reading on the LCD display. 
7.  If there is any question to the accuracy of the reading, perform the  
     "Pure Oxygen Verification Procedure" and if necessary the “Quick  
     One Step Pure Oxygen Calibration Procedure on 

page 6

 

 

CAUTION: 

Do not

 check concentrator with humidifier in place.  

Humidity can affect unit's readings and damage the unit. 

 

 

NOTES: 

Never check a concentrator while holding down the

 "pure O

2

 /calibrate"

button. You 

will get a false reading. You must hold down the 

"pure O

2

 

/calibrate"

 button to check unit 

against a source of U.S.P. 

oxygen (> 99.0%) cylinder or liquid oxygen.

   

 

 

 

 

4

 

Ultrasonic Oxygen Indicator 

 
 

 

 

INSTRUCTIONS FOR SPOT CHECKING 
LIQUID OXYGEN VESSELS or OXYGEN CYLINDERS

 

 
The 

Respond 

O

2

™ 

Ultrasonic Oxygen Indicator

 can also be used to 

perform spot checks (± 2%) of gaseous or liquid oxygen as follows:

 

  

1.  Connect the inlet port on the

 

unit to a gaseous or liquid oxygen 

source (cylinder or liquid oxygen) via a supply tube.  

 
2.   Turn the oxygen flow on to 2 LPM.  
 
3.   Press the 

“on/off”

 button to turn the unit on.  

 
4.   When unit stabilizes the unit display will read "

Concentrator O

2

    

     91.8" 

(approx concentration.) 

 

5.   Press and continue to hold down the 

“pure O

2

/calibrate”

 button.  

 
6.   Unit will display 

"Pure O2"

 and a numeric reading.  

 
7.   If the source gas is oxygen, the display should read between 

99%  

      and 101%

.  

 
8.   If the unit display reads 

"err"

"---",

 

"-U-"

 or any other reading  

other than 99.9 (± 2%), the source gas may not be pure oxygen   
and should be quarantined for retesting using a more sophisticated 
method such as a Servomex* oxygen analyzer.  

 

NOTE: 

Respond 

O

2

™ 

Ultrasonic Oxygen Indicator

 can be used to spot check 

the identity and purity of gaseous or liquid oxygen only, it cannot be used to 
perform or verify U.S.P. oxygen (> 99.0%) testing since it cannot provide the 
required U.S.P. oxygen (> 99.0%) accuracy of ± 0.1%.  

 
 

* Servomex Inc. - Sugarland, TX 

 
 
 
 
 
 
 
 
 

Ultrasonic Oxygen Indicator 

Summary of Contents for Respond O2

Page 1: ...e 866 333 4030 Fax 866 333 4035 sales respondo2 com www respondo2 com 803350 Rev B October 2020 Respond O2 Ultrasonic Oxygen Indicator Operation Manual Responsive Respiratory Inc 6244 Lemay Ferry Rd S...

Page 2: ...turn the unit to Responsive Respiratory Responsive Respiratory will be responsible for normal shipping charges for the return of the unit to purchaser All warranties are voided if the Respond O2 unit...

Page 3: ...ration from any source other than a conventional oxygen concentrator using molecular sieve beds This unit will not withstand fluid spillage The unit will not withstand mechanical shock or vibration NO...

Page 4: ...ies Clean battery terminals When unit is turned on display flashes PRO O2 CAL and then turns off Unit requires calibration Perform Full Calibration Procedure outlined in Operation Manual Display reads...

Page 5: ...of an oxygen concentrator adjusting for the argon that passes through the sieve beds 2 When using the unit with the pure O2 calibrate button depressed the software measures oxygen not adjusting for th...

Page 6: ...s Will the unit lose calibration when I change batteries No The unit holds calibration even when the battery is removed 12 Ultrasonic Oxygen Indicator FREQUENTLY ASKED QUESTIONS The following are some...

Page 7: ...t supply tubing to the concentrator oxygen outlet and the inlet port 5 Turn on the device See above 6 The indicator will stabilize in about 10 seconds and provide an accurate reading on the LCD displa...

Page 8: ...ircuit will compensate for variations in temperature The temperature during storage and shipping should not reach below 29 F 34 C or above 160 F 71 C If the unit has been stored at other than normal r...

Page 9: ...cator FULL TWO STEP CALIBRATION PROCEDURE The following procedure may be used to calibrate the indicator to both room air and U S P oxygen 99 0 1 Do not connect unit to oxygen source or introduce oxyg...

Page 10: ...the rear of the unit away from the case Connect the 9 volt battery to the standard 9 volt battery connection and lay the battery sideways in the battery compartment Insert battery door by sliding door...

Reviews: