VERIFICATION  & CALIBRATION INSTRUCTIONS 

 

PURE OXYGEN VERIFICATION PROCEDURE 

To verify that the indicator unit is accurately measuring the concentration of 
oxygen being produced by a concentrator, perform the following steps:  

 

1.  Connect the inlet port on the unit to a USP oxygen source (cylinder or 

liquid oxygen) via a supply tube.  

2.  Turn the O

 

2

 delivery flow to 2 LPM.  

3.  Press the 

“on/off”

 button to turn unit on.  

4.  When unit stabilizes the display will read - 

"Concentrator O

  

2

 91.8"

 

(approx.concentration.). 

 

 

NOTE: When in the Concentrator Mode, the electronic components of the 
indicator automatically compensate for the argon

 

found in the gas produced by an 

oxygen concentrator. As a result the display will read 

90% 

(± 2%) unless this 

function is over-ridden by pushing the "

pure O

 

2

/calibrate

" button). 

 

 

5.  Press and continue to hold down the 

“pure O

 

2

/calibrate” 

button.  

6.  Unit will display 

“Pure O

 

2

 and a numeric reading.  

7.  If the unit is properly calibrated, the display should read between  
     

99% and 101%

. The 

Respond 

O

2

is ready to use.  

8.  If the unit display reads 

"err, "---", "-U-"

 or any other reading  

other than 99.9 (± 2%), the unit should be recalibrated (See below).  

 

QUICK “ONE STEP” PURE OXYGEN CALIBRATION

 

The following procedure can be used to calibrate the

 Ultrasonic Oxygen 

Concentrator Indicator 

to pure oxygen: 

 

1.  Follow steps 1 through 6 of "Pure Oxygen Verification Procedure"  
     above.  
2.  When the unit displays 

"Pure O

 

2

"

 and a numeric reading,  

continue holding down the 

“pure O

 

2

/calibrate”

 button.  

3.  After a few seconds, the LCD display (while continuing to read    

     "Pure O

2

"

) will start to alternately display 

"CAL"

 and number.  

4.  Continue to hold the 

“pure O

 

2

/calibrate”

 button down until  

     the display stops alternating and simply displays 

“Pure O

 

2

 100”

5.  At that point, the unit is calibrated to pure oxygen and ready for  

use.  

 

NOTE: If unit displays an 

O

 

2

 concentration reading outside of the ± 2% tolerance range 

(eg. < 98% or >102%) unit will not “Quick Cal” and the Full Two Step procedure on 
the following page must be followed. 

 

  
 

 

     

 

 

 

 

 

 

 

 6

 

Ultrasonic Oxygen Indicator 

 
 

 

FULL "TWO STEP" CALIBRATION PROCEDURE

 

The following procedure may be used to calibrate the indicator to both room air 
and U.S.P. oxygen (> 99.0%):  
1. 

  Do not

 connect unit to oxygen source or introduce oxygen to the  

unit at this time.  

2.    Turn unit on by pushing 

“on/off”

 button.  

3.    Wait for the unit to stabilize at room air - and the LCD displays  

       “Concentrator O

 

2

 20.9%”. 

 

If necessary, airflow will purge the   

       

oxygen.  Do not continue to flow air during the rest of this 

       calibration process. 

 

NOTE: Do not exhale breath into unit. Moisture will affect calibration; a slight 
suction can be used to evacuate the sampling chamber. 

 

4.  When unit is stabilized at room air and displays 

“Concentrator  

O

 

2

  20.9%”

 (± 2%) - turn unit OFF by pushing 

“on/off”

 button.

 

5.    Next, hold the 

“pure O

 

2

/calibrate” 

button down while pressing   

       and releasing the 

“on/off”

 button.  

6.    Continue to hold down the 

“pure O

  

2

/calibrate”

 button until the  

       display reads 

“---”

7.    Release the 

“pure O

  

2

/calibrate”

 button and the display will read 

       

“CAL AIR”

.

  

8.    Press and release the 

“pure O

  

2

/calibrate”

 button again 

9.    The unit's display will roll for approximately one minute while it         
       calibrates to room air.  
10.  When the unit display reads 

“CAL O

  

2

 - connect the inlet port  

 

   to an oxygen source (cylinder or liquid) using the supplied tube. 

11.  Introduce U.S.P. oxygen (> 99.0%) at 2 LPM to the unit.  

(Do not  

 

  use  a humidifier). 

12.  Press and release the 

“pure O

2

/calibrate”

 button again. 

13.  The unit's display will roll for approximately one minute while it  
       calibrates to oxygen. 
14.  When the calibration sequence is complete the unit will display 
       

“CAL END”

.  

15.  The unit is then calibrated and ready for use. 
16.  To verify that the unit is properly calibrated, repeat the "Pure Oxygen  
       Verification Procedure” on 

page 6

.  

 

NOTE: 

If unit displays 

"CAL ERR"

 during verification or calibration  

procedures, check to ensure that you are using U.S.P. oxygen (> 99.0%) and that the 
flow is turned on to at least 2 LPM, then repeat procedure. 

 

NOTE: If unit displays "---" during verification or calibration procedures, release the 

"pure 

O2/calibrate”

 button or the unit will shut off automatically. 

 
If any problems are encountered during the calibration sequences,  
contact Responsive Respiratory at 1-866-333-4030. 

 

7

 

Ultrasonic Oxygen Indicator 

Summary of Contents for Respond O2

Page 1: ...e 866 333 4030 Fax 866 333 4035 sales respondo2 com www respondo2 com 803350 Rev B October 2020 Respond O2 Ultrasonic Oxygen Indicator Operation Manual Responsive Respiratory Inc 6244 Lemay Ferry Rd S...

Page 2: ...turn the unit to Responsive Respiratory Responsive Respiratory will be responsible for normal shipping charges for the return of the unit to purchaser All warranties are voided if the Respond O2 unit...

Page 3: ...ration from any source other than a conventional oxygen concentrator using molecular sieve beds This unit will not withstand fluid spillage The unit will not withstand mechanical shock or vibration NO...

Page 4: ...ies Clean battery terminals When unit is turned on display flashes PRO O2 CAL and then turns off Unit requires calibration Perform Full Calibration Procedure outlined in Operation Manual Display reads...

Page 5: ...of an oxygen concentrator adjusting for the argon that passes through the sieve beds 2 When using the unit with the pure O2 calibrate button depressed the software measures oxygen not adjusting for th...

Page 6: ...s Will the unit lose calibration when I change batteries No The unit holds calibration even when the battery is removed 12 Ultrasonic Oxygen Indicator FREQUENTLY ASKED QUESTIONS The following are some...

Page 7: ...t supply tubing to the concentrator oxygen outlet and the inlet port 5 Turn on the device See above 6 The indicator will stabilize in about 10 seconds and provide an accurate reading on the LCD displa...

Page 8: ...ircuit will compensate for variations in temperature The temperature during storage and shipping should not reach below 29 F 34 C or above 160 F 71 C If the unit has been stored at other than normal r...

Page 9: ...cator FULL TWO STEP CALIBRATION PROCEDURE The following procedure may be used to calibrate the indicator to both room air and U S P oxygen 99 0 1 Do not connect unit to oxygen source or introduce oxyg...

Page 10: ...the rear of the unit away from the case Connect the 9 volt battery to the standard 9 volt battery connection and lay the battery sideways in the battery compartment Insert battery door by sliding door...