E-21
Helpful hints
When the red check lamp flashes, you
need to check the cell and power supply
itself. When the green working lamp is
lighten up, it shows the power supply is
working correctly.
POWER SUPPLY
1. Check and
working lamp
3. Power supply
ground wire
connection.
Ground wire connection for power
supply and electronic cell
Power supply power
cord from transformer
High voltage
output adjustor
High voltage
output adjustor
Check and
working lamp
2. Red twin check
wire connection
The power supply is one of the most important parts of the unit. The power supply provides high voltages to the cell through
the power contact plates. The red wire twin check wire connect the main circuit board. The main circuit board always
investigate the voltage on the check wire, if the voltage is less than 1V dc, the main circuit board will light up the CHECK
indicator light on the control panel so that you can fix the problem as soon as possible.
ENGLISH
PART III
IMPORTANT PARTS - POWER SUPPLY AND TRANSFORMER