1
Unplug the power cord from the power outlet
(G-1)
.
!
WARNING
Take care when handling your HumidAire 3i. The water and the water chamber
base may both be hot. Allow ten minutes for the heater plate and any excess water
to cool (G-2).
2
Remove the water chamber from the docking station
(G-3)
.
3
Press the
large silver button
located midway
underneath
the docking station
(G-4)
and pull the
H
UMID
A
IRE
3i
and flow generator apart
(G-5)
.
Note
: Re-attach the connector plug firmly to your flow generator.
F
REQUENTLY
A
SKED
Q
UESTIONS
W
HY
IS
THERE
CONDENSATION
IN
MY
MASK
AND
/
OR
AIR
TUBING
?
You may experience excessive condensation in your mask or air tubing. This may occur
seasonally, particularly when the room temperature is low or cool air is circulating
around the tube. Turn the control dial
down
to a lower setting until the condensation
decreases.
You may have to adjust your humidity setting from season to season.
D
O
I
HAVE
TO
USE
DISTILLED
WATER
IN
THE
WATER
CHAMBER
?
Using distilled water will maximise the life of the water chamber and reduce mineral
deposits. See “Cleaning your HumidAire 3i” on page 10 for further details.
W
HAT
IF
I
STILL
HAVE
SYMPTOMS
OF
DRYNESS
?
If you still experience symptoms of dryness, then your humidity setting may be set too
low to be effective. Gradually turn the humidity control dial up over a few nights until
the symptoms disappear.
If you are still experiencing symptoms of dryness, even after increasing the humidity
setting, you may be breathing through your mouth. Mouth breathing or mouth leaks
cause drying and irritation of the upper airway. If this is a problem, a full face mask or
chin strap may help. Consult your clinician or equipment supplier for details.